Best viewed on a computer. These pages are not yet optimized for phones: some figures and controls may not fit the screen.
Basics Advanced
Learn · for people who already use a microscope

Microscopy advanced

The physics behind the settings you choose at the microscope.

Ten modules, from the photon to the pixel. Every plot is computed from a stated formula or a vectorial optical model, every simulated image says so, and every number has a source.

10 modules2 calculators31 cited sources
Computed side view (x–z) of the image of one point in a widefield microscope: an hourglass, 209 by 510 nanometres at half maximum, whose light spreads out above and below focusWidefield
lateral × axial FWHM209 × 510 nm
Computed side view (x–z) of the image of one point in a confocal at 1 Airy unit: compact, 187 by 391 nanometresConfocal 1 AU
lateral × axial FWHM187 × 391 nm
Computed side view (x–z) of the image of one point in a confocal at 0.2 Airy units: smaller still, 147 by 355 nanometresConfocal 0.2 AU
lateral × axial FWHM147 × 355 nm
Signal from a thin layer, by distance from focusCOMPUTED00.5100.511.52distance from focus (µm) · y: signal, 1 = in focuswidefield: every layer sends the same lightconfocal 1 AU: section 501 nm thick (FWHM)confocal 0.2 AU: 399 nm
Computed · The image of a single point, seen from the side (x–z, 2.4 × 4 µm per panel), NA 1.4 oil, 488/520 nm, vectorial model. Each panel is scaled to its own peak on a square-root scale, to compare shapes: on one scale the 0.2 AU image would peak at about 0.13 of the 1 AU one (module 10). The widefield point image is short along z too (510 nm), but out of focus its light spreads out instead of disappearing, so each plane still sends the same total light and widefield has no optical sectioning (module 02). The chart shows exactly that: the signal from a thin uniform layer, from the same model, stays at 1 for widefield at every distance and falls for the confocal; its width is the optical section (thin-sheet FWHM)[5].
Start from your question

What are you trying to fix?

Pick the problem. It takes you to the module that computes the answer, with the formula and its source.

live model “My images are hazy in thick samples.” Out-of-focus light reaches the detector. See what the pinhole blocks, and what it costs in signal. also 02 · The confocal principle03 · The pinhole →
live model “Two structures blur into one.” Check the limit for your objective and wavelength, then what pixel reassignment or deconvolution adds. also 10 · Airyscan and LIGHTNING04 · Resolution →
simulated “What pixel size, zoom or pixels per line?” Nyquist by the practical and the bandwidth rule, with a calculator for your objective. 05 · Nyquist →
live model “How many z slices, how far apart?” Section thickness, step and slice count, and why index mismatch stretches a stack. 06 · Z-stacks →
live model “My green dye shows up in the red channel.” Measured spectra, detection bands, sequential scanning and unmixing, with the mixing matrix. 07 · Spectra →
live model “The signal fades during my time-lapse.” Dose per frame, frames to half signal, and what averaging really buys. 08 · Photobleaching →
live model “I need to image deep inside tissue.” Two-photon excitation: depth without a pinhole, at a cost in resolution. 09 · Multiphoton →
live model “Which laser line, and is my dye bright enough?” Spectra, extinction coefficient, quantum yield and the expected signal per line. 01 · Fluorescence →
10 modules · 5 chapters

The modules

Legend

How to read the pictures

computed

A number, curve or image calculated from a stated physical model (point spread functions, spectra, transfer functions), drawn to scale.

simulated

A synthetic sample passed through a model of the optics, with or without photon noise. It looks like data; it is not.

model

An interactive or illustrative model with stated simplifications (for example first-order bleaching). “Live” means it recomputes as you change the inputs.

render

A 3D illustration of a generic instrument, with proportions adjusted so the light path reads. An illustration, never evidence for a number.

diagram · not to scale

A drawing that cannot be to scale. The fine print says what is not to scale.

New to microscopy?The same ideas, in pictures and plain words: Microscopy basics. Go to Microscopy basics →

References

Shared by all ten modules, in one numbered list. Figures from a manufacturer rather than the literature are identified as such where they are used.

  1. [1]Lakowicz JR. Principles of Fluorescence Spectroscopy, 3rd ed. Springer, 2006. Chapter 1 (Jablonski diagram, Stokes shift, quantum yield, lifetime).
  2. [2]Minsky M. Microscopy apparatus. US Patent 3,013,467, filed 7 November 1957, issued 19 December 1961.
  3. [3]Born M, Wolf E. Principles of Optics, 6th ed. Pergamon, 1980. §8.5.2 (Airy pattern, encircled energy), §8.6.2 (Rayleigh criterion), §8.6.3 (image formation in the microscope: incoherent illumination; coherent illumination, Abbe’s theory) and §8.8 (three-dimensional light distribution near focus).
  4. [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
  5. [5]Wilson T. Resolution and optical sectioning in the confocal microscope. Journal of Microscopy 244:113–121, 2011. doi:10.1111/j.1365-2818.2011.03549.x
  6. [6]Sheppard CJR, Matthews HJ. Imaging in high-aperture optical systems. Journal of the Optical Society of America A 4:1354–1360, 1987.
  7. [7]Abbe E. Beiträge zur Theorie des Mikroskops und der mikroskopischen Wahrnehmung. Archiv für mikroskopische Anatomie 9:413–468, 1873.
  8. [8]Rayleigh, Lord. Investigations in optics, with special reference to the spectroscope. Philosophical Magazine (5) 8:261–274, 1879.
  9. [9]Pawley JB (ed.). Handbook of Biological Confocal Microscopy, 3rd ed. Springer, 2006. Chapter 4 (Pawley, “Points, pixels, and gray levels”) and Chapter 39 (Diaspro et al., “Photobleaching”).
  10. [10]Scientific Volume Imaging. “Nyquist rate” and “Nyquist calculator”, svi.nl.
  11. [11]Zimmermann T. Spectral imaging and linear unmixing in light microscopy. Advances in Biochemical Engineering/Biotechnology 95:245–265, 2005. doi:10.1007/b102216
  12. [12]Song L, Hennink EJ, Young IT, Tanke HJ. Photobleaching kinetics of fluorescein in quantitative fluorescence microscopy. Biophysical Journal 68:2588–2600, 1995. doi:10.1016/S0006-3495(95)80442-X
  13. [13]Hell S, Reiner G, Cremer C, Stelzer EHK. Aberrations in confocal fluorescence microscopy induced by mismatches in refractive index. Journal of Microscopy 169:391–405, 1993.
  14. [14]Denk W, Strickler JH, Webb WW. Two-photon laser scanning fluorescence microscopy. Science 248:73–76, 1990. doi:10.1126/science.2321027
  15. [15]Helmchen F, Denk W. Deep tissue two-photon microscopy. Nature Methods 2:932–940, 2005. doi:10.1038/nmeth818
  16. [16]Sheppard CJR. Super-resolution in confocal imaging. Optik 80:53–54, 1988; Müller CB, Enderlein J. Image scanning microscopy. Physical Review Letters 104:198101, 2010.
  17. [17]Huff J. The Airyscan detector from ZEISS: confocal imaging with improved signal-to-noise ratio and super-resolution. Nature Methods 12:i–ii, 2015 (manufacturer application note).
  18. [18]Zipfel WR, Williams RM, Webb WW. Nonlinear magic: multiphoton microscopy in the biosciences. Nature Biotechnology 21:1369–1377, 2003. doi:10.1038/nbt899
  19. [19]Schermelleh L, Ferrand A, Huser T, Eggeling C, Sauer M, Biehlmaier O, Drummen GPC. Super-resolution microscopy demystified. Nature Cell Biology 21:72–84, 2019. doi:10.1038/s41556-018-0251-8
  20. [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.
  21. [21]Lambert TJ. FPbase: a community-editable fluorescent protein database. Nature Methods 16:277–278, 2019. Spectra from fpbase.org (CC BY-SA 4.0). doi:10.1038/s41592-019-0352-8
  22. [22]Laissue PP, Alghamdi RA, Tomancak P, Reynaud EG, Shroff H. Assessing phototoxicity in live fluorescence imaging. Nature Methods 14:657–661, 2017. doi:10.1038/nmeth.4344
  23. [23]Diel EE, Lichtman JW, Richardson DS. Tutorial: avoiding and correcting sample-induced spherical aberration artifacts in 3D fluorescence microscopy. Nature Protocols 15:2773–2784, 2020. doi:10.1038/s41596-020-0360-2
  24. [24]Richardson WH. Bayesian-based iterative method of image restoration. Journal of the Optical Society of America 62:55–59, 1972; Lucy LB. An iterative technique for the rectification of observed distributions. Astronomical Journal 79:745–754, 1974.
  25. [25]Goodman JW. Introduction to Fourier Optics, 4th ed. W. H. Freeman, 2017. §7.5 (comparison of coherent and incoherent imaging; §7.5.2, two-point resolution).
  26. [26]Enderlein J. Theoretical study of detection of a dipole emitter through an objective with high numerical aperture. Optics Letters 25:634–636, 2000.
  27. [27]Shaner NC, Steinbach PA, Tsien RY. A guide to choosing fluorescent proteins. Nature Methods 2:905–909, 2005. doi:10.1038/nmeth819
  28. [28]Eggeling C, Widengren J, Rigler R, Seidel CAM. Photobleaching of fluorescent dyes under conditions used for single-molecule detection: evidence of two-step photolysis. Analytical Chemistry 70:2651–2659, 1998. doi:10.1021/ac980027p
  29. [29]Donnert G, Eggeling C, Hell SW. Major signal increase in fluorescence microscopy through dark-state relaxation. Nature Methods 4:81–86, 2007. doi:10.1038/nmeth986
  30. [30]Patterson GH, Piston DW. Photobleaching in two-photon excitation microscopy. Biophysical Journal 78:2159–2162, 2000. doi:10.1016/S0006-3495(00)76762-2
  31. [31]Coltman JW. The specification of imaging properties by response to a sine wave input. Journal of the Optical Society of America 44:468–471, 1954. doi:10.1364/JOSA.44.000468

Training

These pages are purely to improve your knowledge: practical training on the instrument is not only recommended but mandatory.