Module 08 · Chapter IV · Signal and samples

Photobleaching: a finite photon budget

The dye’s photon budget, how power, averaging and frame count divide it, what a z-stack costs every plane, and why phototoxicity needs its own check.

SIMULATED · Two ways to spend the same dye. Top: 50 frames at 4× power; bottom: 200 frames at 1×. Both bleach 75 % of the dye and collect about 1,090 photons per bright pixel; the top series puts them into fewer, cleaner frames. Each image is scaled to its expected brightness, so only the grain differs (first-order model, section 2).
1 · The physics

In the linear regime, dose sets the bleaching

Every fluorophore survives only a limited number of excitation cycles before a photochemical reaction, often from the triplet state and often involving oxygen, destroys its ability to fluoresce. The number of photons a molecule emits before it bleaches is roughly the ratio of its fluorescence and bleaching quantum yields, Φf/Φb. For fluorescein in an oxygen-rich medium the probability of bleaching is about 3 × 10⁻⁵ per excitation, so a molecule emits roughly 10⁴ to 3 × 10⁴ photons before it bleaches (Zeiss monograph, Part 3, p. IV[4]; Shaner et al. 2005[27]). EGFP emits roughly 24 to 33 times more under lamp illumination (Shaner et al. 2005). Under focused laser light both numbers can be much lower, because at high intensity bleaching can grow faster than linearly (Eggeling et al. 1998[28]; Donnert et al. 2007[29]).

In the linear regime, bleaching scales with the cumulative excitation dose, intensity × time. The simplest description is first-order decay, F(t) = F0 e−kt with k proportional to intensity; for fluorescein in thin samples the one-photon bleaching rate was measured to grow in proportion to the power (Patterson and Piston 2000[30]). It is a useful planning tool, but measured kinetics often deviate from a single exponential and depend on oxygen, mounting medium and antifade.[12]

With one-photon excitation the laser lights every plane it passes through, so recording one plane bleaches the whole thickness the cone reaches, and a z-stack bleaches every plane once per slice (figure). Light also damages living cells, frequently before any fading is visible. Phototoxicity should be assessed on the biology itself (division, motility, morphology) at the imaging settings, not inferred from bleaching alone.[22] How the dose is delivered matters too: in one study, spacing the excitation into pulses more than 1 µs apart, so that dark (triplet) states could relax, gave 5 to 25 times more photons before bleaching than strong continuous or high-repetition-rate pulsed light, such as a white light laser at 80 MHz (Donnert et al. 2007).

Slices in the stack39 slices · 9.50 µm deep
Slices 0.25 µm apart, centred on the focus
Stack presets
Dose per slice1×
Light dose per area per slice, relative to the reference frame

39 slices at 1×: every plane in reach of the cone has had 39 frame-doses of light and lost 24% of its dye before the stack is finished: the last slices are dimmer than the first, and so is the next stack.

COMPUTED · Light map: the vectorial point spread function of the focused 488 nm beam at NA 1.4 oil (circular polarisation, back aperture filled evenly), log scale over four decades, drawn to scale; the dashed lines mark the 67.3° half-angle; the brackets mark the radius of the lit footprint. Each plane’s total light per frame equals the focal plane’s (energy conservation; computed: within 0.13 % out to 3 µm). While the laser scans, every plane whose lit footprint (radius 2.39 µm per µm from the focus) stays inside the field receives the same light per area as the focal plane. Ledger: slices 0.25 µm apart (module 06’s “about 250 nm, half the optical section”), the dose per area at the centre of a 50 µm field, where a slice still lights a plane fully up to 10.5 µm away (geometric footprint of the evenly filled aperture); bleaching per plane uses the first-order model of section 2 (half after 100 frame-doses), with the same laser power for every slice; bars on a log scale. Scattering and absorption, which shield deep planes in real tissue, are not modelled. The map assumes a sample index-matched to the oil (n = 1.518); in an aqueous sample the rays beyond NA 1.33 are reflected at the coverslip and the rest spread into a wider cone. Equal light per plane means equal bleaching only in the linear regime; where bleaching grows faster than linearly with intensity, the focal plane loses more. Two-photon comparison (button): the objective switches to NA 1.0 water, as in module 09; the map is the squared intensity of the 976 nm focus and the bars are two-photon excitation per plane, summed over the slices in focal-frame equivalents (computed to 3 µm, beyond that the far-field 1/z² law), at the same power; they are excitation, not bleaching, so no dye loss is computed. In the focus, two-photon bleaching grows even faster than the excitation (rate ∝ power³ or more; Patterson and Piston 2000), so it is confined to the focus but can be stronger there.
2 · Live model

Spend a fixed photon budget

The dye in a bright pixel can deliver only so many detected photons before it is gone. Choose laser power, line averaging and the number of frames: the charts show how fast you spend the budget and where your setting sits among all settings, and the images show your first and last frame.

The dye’s budget: photons collected so far, against what the dye can give · MODEL the dye’s budget: about 1,450 photons per bright pixel you (green dot), frame 200: 1,086 photons · 75% of the dye 02004006008001,00005001,0001,5002,000 frame numberphotons per bright pixel, all frames so far The trade-off: frames against signal-to-noise per frame; each orange line is one share of the dye bleached · MODEL 25%50%75%90%dyebleached you (white dot): 200 frames, SNR 3.2, 75% bleached 1020501002005001,0001251020 frames in the series (log scale)SNR of frame 1 (√photons, log)
First and last frame at your settings · SIMULATED · LIVE · 19.2 × 16 µm
Same dye, three ways (75 % of it each)
Laser power1×
Half-stops from 0.25× to 4× the reference
Line averages
Frames200 frames
Log scale, 10 to 1,000 frames
Frame presets
Dose per frame (× reference)1×
Frame at which half remainsframe 101
Last frame, of the first25%
Photons per bright pixel, first → last10 → 2.5
Signal-to-noise, first → last3.2 → 1.6
Dye used · photons collected75% · 1,086

Most of the dye is gone: the last frame records only 25% of the first (2.5 photons per pixel, mostly grain). This series asks for more than the dye can give: take fewer frames. Lowering the power would leave every frame mostly grain, and more light would only fade it faster.

MODEL · First-order bleaching: frame n records Sn = S1 · 2−(n−1)·d/100, where d is the light dose per area per frame relative to the reference (power × line averages × (80 nm / pixel size)²; pixels are 80 nm here); at the reference (80 nm pixels, 4 µs per pixel, 1×, no averaging) frame 101 records half of frame 1. Same model and constants as lesson 8 and the Basics explorer. Photons: 10 per pixel in frame 1 at the reference, for a bright structure that fills the pixel; photons scale with power × averages. Signal-to-noise = √(photons): photon (Poisson) noise only, as with photon counting; background, detector read noise and the excess noise of analog detectors are not included. Budget: summing the decaying frames gives about 1,450 detected photons per bright pixel for any power or averaging (10 × 100 / ln 2 = 1,443 in the limit; 1,444 to 1,463 from 0.25× to 4×, more at very high doses, where one frame bleaches several percent and counting each frame at its starting brightness overstates the total), so photons collected = budget × share of the dye bleached; its size follows from the two illustrative settings (10 photons, half-life 100 frames), its constancy from the linear model. The verdict’s thresholds (the last frame below 30 %, 50 % or above 90 % of the first) and the grain words (“mostly grain” below 4 photons per pixel, “grainy” below 15) are this site’s conventions, as in the Basics explorer and lesson 8. Images: the simulated cell (confocal, 1 AU, NA 1.4 oil) at 80 nm pixels with photon noise from a fixed seed; both labels fade at the same rate here; “same as frame 1” shows the fading, “each frame scaled” the growing grain. The half-life of 100 frames is illustrative. Real kinetics are often not single-exponential and depend on oxygen and the dye’s surroundings[12]; at the high peak intensities of a point scanner, many dyes bleach faster than linearly and dark (triplet) states build up, so the budget shrinks at high power and grows when dark states are allowed to relax (5–25× more photons in Donnert et al. 2007; Eggeling et al. 1998). Phototoxicity is not modelled and can occur without visible bleaching[22].

Formulas used in this module

The math, with its sources

First-order bleaching[9][12]
F(n) = F1 · 2−(n−1)d/100
n: frame number; d: dose per area per frame relative to the reference, power × dwell time × averages ÷ pixel area (Pawley ch. 4, Fig. 4.8). At the reference, frame 101 records half of frame 1; half remains at frame 1 + 100/d.
Photons before bleaching[4][9]
Nph ≈ Φf / Φb
Φf: fluorescence quantum yield; Φb: probability of bleaching per excitation. Fluorescein in an oxygen-rich medium: Φb ≈ 2.7 × 10⁻⁵; with Φf = 0.71 (the value in [4]) about 26,000 photons emitted, about 3 × 10⁴ for Φf ≈ 0.9.
Shot-noise-limited signal-to-noise[9]
SNR = √N
N: detected photons per pixel. Averaging A lines or frames multiplies N, and the dose, by A.
Photons collected by a series
C = Ntot (1 − 2−N d/100),Ntot = N1 / (1 − 2−d/100)
N: number of frames; N1: photons per bright pixel in frame 1; Ntot: the dye’s budget, about 1,450 per bright pixel in this model at any power or averaging. The bracket is the share of the dye bleached.
The trade-off
N · SNR1² = 1000 · log2(1/(1 − b))
Frames × (signal-to-noise of frame 1)² is fixed by the share b of the dye you bleach: 1,000 at 50 %, 2,000 at 75 %, 3,322 at 90 %. Each 1 % of dye buys about 14.4 photons per bright pixel (80 nm pixels; × (pixel/80 nm)² for others). This model.

About the model

First-order decay in the accumulated dose (rate constant proportional to intensity) holds only in the linear excitation regime. At high intensities, molecules that are already excited absorb again and break down from higher singlet or triplet states (two-step photolysis; Eggeling et al. 1998[28]), so damage grows faster than linearly and each molecule gives fewer photons before it bleaches. Saturation, separately, makes the signal grow slower than linearly with power. Oxygen depletion changes the rate over time.[12] Saturation alone already bends the curve: for EGFP at NA 1.4, singlet saturation costs about 3 % of the signal at the centre of the focus at 10 µW at the sample and about 26 % at 100 µW (paraxial estimate for a continuous-wave laser; a pulsed white-light laser at 80 MHz saturates about twice as fast; the image as a whole loses less, and triplet build-up lowers the threshold).

At the microscope

Spend the budget where it counts: the lowest dose that answers the question, no more frames or slices than needed, and Nyquist sampling rather than oversampling (module 05).

For live cells, check viability or behavior at your settings; bleaching is not a safe proxy for phototoxicity.[22]

Check yourself

Four line averages at 1× power, or one scan at 4× power: what changes in the linear regime?

References

Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.

  1. [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
  2. [9]Pawley JB (ed.). Handbook of Biological Confocal Microscopy, 3rd ed. Springer, 2006. Chapter 4 (Pawley, “Points, pixels, and gray levels”) and Chapter 39 (Diaspro et al., “Photobleaching”).
  3. [12]Song L, Hennink EJ, Young IT, Tanke HJ. Photobleaching kinetics of fluorescein in quantitative fluorescence microscopy. Biophysical Journal 68:2588–2600, 1995. doi:10.1016/S0006-3495(95)80442-X
  4. [22]Laissue PP, Alghamdi RA, Tomancak P, Reynaud EG, Shroff H. Assessing phototoxicity in live fluorescence imaging. Nature Methods 14:657–661, 2017. doi:10.1038/nmeth.4344
  5. [27]Shaner NC, Steinbach PA, Tsien RY. A guide to choosing fluorescent proteins. Nature Methods 2:905–909, 2005. doi:10.1038/nmeth819
  6. [28]Eggeling C, Widengren J, Rigler R, Seidel CAM. Photobleaching of fluorescent dyes under conditions used for single-molecule detection: evidence of two-step photolysis. Analytical Chemistry 70:2651–2659, 1998. doi:10.1021/ac980027p
  7. [29]Donnert G, Eggeling C, Hell SW. Major signal increase in fluorescence microscopy through dark-state relaxation. Nature Methods 4:81–86, 2007. doi:10.1038/nmeth986
  8. [30]Patterson GH, Piston DW. Photobleaching in two-photon excitation microscopy. Biophysical Journal 78:2159–2162, 2000. doi:10.1016/S0006-3495(00)76762-2