Photobleaching: a finite photon budget
The dye’s photon budget, how power, averaging and frame count divide it, what a z-stack costs every plane, and why phototoxicity needs its own check.
In the linear regime, dose sets the bleaching
Every fluorophore survives only a limited number of excitation cycles before a photochemical reaction, often from the triplet state and often involving oxygen, destroys its ability to fluoresce. The number of photons a molecule emits before it bleaches is roughly the ratio of its fluorescence and bleaching quantum yields, Φf/Φb. For fluorescein in an oxygen-rich medium the probability of bleaching is about 3 × 10⁻⁵ per excitation, so a molecule emits roughly 10⁴ to 3 × 10⁴ photons before it bleaches (Zeiss monograph, Part 3, p. IV[4]; Shaner et al. 2005[27]). EGFP emits roughly 24 to 33 times more under lamp illumination (Shaner et al. 2005). Under focused laser light both numbers can be much lower, because at high intensity bleaching can grow faster than linearly (Eggeling et al. 1998[28]; Donnert et al. 2007[29]).
In the linear regime, bleaching scales with the cumulative excitation dose, intensity × time. The simplest description is first-order decay, F(t) = F0 e−kt with k proportional to intensity; for fluorescein in thin samples the one-photon bleaching rate was measured to grow in proportion to the power (Patterson and Piston 2000[30]). It is a useful planning tool, but measured kinetics often deviate from a single exponential and depend on oxygen, mounting medium and antifade.[12]
With one-photon excitation the laser lights every plane it passes through, so recording one plane bleaches the whole thickness the cone reaches, and a z-stack bleaches every plane once per slice (figure). Light also damages living cells, frequently before any fading is visible. Phototoxicity should be assessed on the biology itself (division, motility, morphology) at the imaging settings, not inferred from bleaching alone.[22] How the dose is delivered matters too: in one study, spacing the excitation into pulses more than 1 µs apart, so that dark (triplet) states could relax, gave 5 to 25 times more photons before bleaching than strong continuous or high-repetition-rate pulsed light, such as a white light laser at 80 MHz (Donnert et al. 2007).
39 slices at 1×: every plane in reach of the cone has had 39 frame-doses of light and lost 24% of its dye before the stack is finished: the last slices are dimmer than the first, and so is the next stack.
Spend a fixed photon budget
The dye in a bright pixel can deliver only so many detected photons before it is gone. Choose laser power, line averaging and the number of frames: the charts show how fast you spend the budget and where your setting sits among all settings, and the images show your first and last frame.
Most of the dye is gone: the last frame records only 25% of the first (2.5 photons per pixel, mostly grain). This series asks for more than the dye can give: take fewer frames. Lowering the power would leave every frame mostly grain, and more light would only fade it faster.
MODEL · First-order bleaching: frame n records Sn = S1 · 2−(n−1)·d/100, where d is the light dose per area per frame relative to the reference (power × line averages × (80 nm / pixel size)²; pixels are 80 nm here); at the reference (80 nm pixels, 4 µs per pixel, 1×, no averaging) frame 101 records half of frame 1. Same model and constants as lesson 8 and the Basics explorer. Photons: 10 per pixel in frame 1 at the reference, for a bright structure that fills the pixel; photons scale with power × averages. Signal-to-noise = √(photons): photon (Poisson) noise only, as with photon counting; background, detector read noise and the excess noise of analog detectors are not included. Budget: summing the decaying frames gives about 1,450 detected photons per bright pixel for any power or averaging (10 × 100 / ln 2 = 1,443 in the limit; 1,444 to 1,463 from 0.25× to 4×, more at very high doses, where one frame bleaches several percent and counting each frame at its starting brightness overstates the total), so photons collected = budget × share of the dye bleached; its size follows from the two illustrative settings (10 photons, half-life 100 frames), its constancy from the linear model. The verdict’s thresholds (the last frame below 30 %, 50 % or above 90 % of the first) and the grain words (“mostly grain” below 4 photons per pixel, “grainy” below 15) are this site’s conventions, as in the Basics explorer and lesson 8. Images: the simulated cell (confocal, 1 AU, NA 1.4 oil) at 80 nm pixels with photon noise from a fixed seed; both labels fade at the same rate here; “same as frame 1” shows the fading, “each frame scaled” the growing grain. The half-life of 100 frames is illustrative. Real kinetics are often not single-exponential and depend on oxygen and the dye’s surroundings[12]; at the high peak intensities of a point scanner, many dyes bleach faster than linearly and dark (triplet) states build up, so the budget shrinks at high power and grows when dark states are allowed to relax (5–25× more photons in Donnert et al. 2007; Eggeling et al. 1998). Phototoxicity is not modelled and can occur without visible bleaching[22].
The math, with its sources
About the model
First-order decay in the accumulated dose (rate constant proportional to intensity) holds only in the linear excitation regime. At high intensities, molecules that are already excited absorb again and break down from higher singlet or triplet states (two-step photolysis; Eggeling et al. 1998[28]), so damage grows faster than linearly and each molecule gives fewer photons before it bleaches. Saturation, separately, makes the signal grow slower than linearly with power. Oxygen depletion changes the rate over time.[12] Saturation alone already bends the curve: for EGFP at NA 1.4, singlet saturation costs about 3 % of the signal at the centre of the focus at 10 µW at the sample and about 26 % at 100 µW (paraxial estimate for a continuous-wave laser; a pulsed white-light laser at 80 MHz saturates about twice as fast; the image as a whole loses less, and triplet build-up lowers the threshold).
At the microscope
Spend the budget where it counts: the lowest dose that answers the question, no more frames or slices than needed, and Nyquist sampling rather than oversampling (module 05).
For live cells, check viability or behavior at your settings; bleaching is not a safe proxy for phototoxicity.[22]
Four line averages at 1× power, or one scan at 4× power: what changes in the linear regime?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
- [9]Pawley JB (ed.). Handbook of Biological Confocal Microscopy, 3rd ed. Springer, 2006. Chapter 4 (Pawley, “Points, pixels, and gray levels”) and Chapter 39 (Diaspro et al., “Photobleaching”).
- [12]Song L, Hennink EJ, Young IT, Tanke HJ. Photobleaching kinetics of fluorescein in quantitative fluorescence microscopy. Biophysical Journal 68:2588–2600, 1995. doi:10.1016/S0006-3495(95)80442-X
- [22]Laissue PP, Alghamdi RA, Tomancak P, Reynaud EG, Shroff H. Assessing phototoxicity in live fluorescence imaging. Nature Methods 14:657–661, 2017. doi:10.1038/nmeth.4344
- [27]Shaner NC, Steinbach PA, Tsien RY. A guide to choosing fluorescent proteins. Nature Methods 2:905–909, 2005. doi:10.1038/nmeth819
- [28]Eggeling C, Widengren J, Rigler R, Seidel CAM. Photobleaching of fluorescent dyes under conditions used for single-molecule detection: evidence of two-step photolysis. Analytical Chemistry 70:2651–2659, 1998. doi:10.1021/ac980027p
- [29]Donnert G, Eggeling C, Hell SW. Major signal increase in fluorescence microscopy through dark-state relaxation. Nature Methods 4:81–86, 2007. doi:10.1038/nmeth986
- [30]Patterson GH, Piston DW. Photobleaching in two-photon excitation microscopy. Biophysical Journal 78:2159–2162, 2000. doi:10.1016/S0006-3495(00)76762-2