Best viewed on a computer. These pages are not yet optimized for phones: some figures and controls may not fit the screen.
Lesson 8 of 8 All lessons
Microscopy basics · Lesson 8 of 8 · about 5 minutes

Light is a budget

Every image fades the dye a little and stresses living cells. Light spent on one image isn’t available for the next.

Trades: Clean image · Sample health Widefield too Go deeper: the advanced module on photobleaching →
Simulated image 1 of the same cell, 100% of the signal of image 1 left, at the same display settingsimage 1100% left
Simulated image 50 of the same cell, 71% of the signal of image 1 left, at the same display settingsimage 5071% left
Simulated image 100 of the same cell, 50% of the signal of image 1 left, at the same display settingsimage 10050% left
Simulated image 200 of the same cell, 25% of the signal of image 1 left, at the same display settingsimage 20025% left
SIMULATED · Images 1, 50, 100 and 200 of one time-lapse series of the same field (48.6 × 32.4 µm), all at the same display settings. Each image fades the dye by the same small fraction: image 101 records half the signal of image 1 (first-order model, as in the Basics explorer; the 100-image half-life is illustrative). Both labels fade at the same rate here for simplicity; real dyes bleach at different rates.
1 · The idea

Dyes don’t last forever

A dye molecule can absorb and give off light many times, but every cycle carries a small chance that it breaks and goes dark for good. This is photobleaching. Depending on the dye, molecules last from tens of thousands to millions of cycles.

Light also makes reactive oxygen molecules that can stress or kill living cells (phototoxicity), often before any fading is visible.

So the light you put into a sample is a budget. A clean image needs many photons; a long time-lapse needs many images; a z-stack needs many slices. All draw on the same account. Imaging well means spending only what your question needs.

Every plane receives the same light per scan, so dye 2 µm above the focus fades as fast as dye in focus; yet the pinhole passes its light at only 0.8% of the rate for the in-focus plane.

COMPUTED · The focused laser beam in a side view (x–z), drawn to scale for a 1.4 NA oil objective: a double cone with a 67.3° half-angle. Every plane inside it receives the same total light per scan; the pinhole only decides what reaches the detector.

Light map: vectorial point spread function (Richards–Wolf) of a 1.4 NA oil objective (n = 1.518) at 488 nm with the back aperture filled evenly, in a 12 × 6 µm side view at 40 nm per pixel, on a logarithmic scale over four decades below the in-focus peak. Light per plane: the power through each plane, equal to within 0.1% from the focus to 2 µm. Recorded signal: a thin uniform dye layer at each depth, pinhole 1 Airy unit (excitation × detection through the pinhole), relative to the same layer in focus. Equal fading in every plane assumes one-photon excitation and fading in proportion to the light (the model of the try-it below); where bleaching grows faster than the light, the focal plane fades somewhat faster. Near the centre of the scanned field (in a 50 µm field, for planes up to about 10 µm from the focus), the light per area equals that in the focal plane. Scattering and absorption are not modelled.

2 · Try it

Spend the budget

Choose a laser power and how many images you need. The chart shows what each image records in a bright pixel; the bar shows how much of the dye’s fixed photon budget your series spends; the images show the cost.

Photons each image records in a bright pixel. The filled area is what your series collects; the dashed line is what is still in the dye. 010203040 0100200300400500 image numberphotons per pixel, bright structure mostly grainbelow 4grainybelow 15 image 200: 2.5 photons MODEL
The dye’s budget: ≈ 1,450 detected photons per bright pixelabout the same however you spend them
Spent by your series: 75% (≈ 1,090 photons)Left in the dye: 25%
Three images of your series, with photon noise, all on the display of image 1 (as on the microscope).
Laser power · relative to the reference setting1×
Presets
Images you need200 images
Presets
Photons per pixel, first → last image10 → 2.5grainy → mostly grain
Last image, of the first25%
Dye used by the series · = share of the budget spent75%

Most of the dye is gone: the last image records only 25% of the first (2.5 photons per pixel, mostly grain). This series asks for more than the dye can give: take fewer images. Lowering the power would leave every image mostly grain, and more light would only fade it faster.

MODEL · First-order bleaching: each image removes the same fraction of the remaining dye, in proportion to the light dose per image (here the laser power). At the reference setting (80 nm pixels, 4 µs per pixel, 1×, no averaging) image 101 records half the signal of image 1: the same model and constants as the Basics explorer and module 08. Photons per pixel: 10 in the first image at the reference setting, for a bright structure that fills the pixel. Because the fading per image grows with the power exactly as the photons do, the dye yields about the same total, about 1,450 detected photons per bright pixel, however you spend it; that total is about 10 photons × 100 images ÷ ln 2, so its size follows from this page’s illustrative settings, while the fact that it is fixed is the general point. Images: the simulated cell (confocal, 1 Airy unit, NA 1.4 oil) at 80 nm pixels with photon noise, all shown at the display of the first image (or each scaled to its own expected brightness, with “Auto-scale each image” on); both labels fade at the same rate here. Signal-to-noise = √(photons), photon noise only; the words “mostly grain” (fewer than 4 photons per pixel) and “grainy” (fewer than 15) are this site’s conventions, as in the Basics explorer. The half-life of 100 images is illustrative, not a property of a particular dye. Bleaching in proportion to the power matches measurements on fluorescein at moderate power (Patterson & Piston 2000, Biophys J 78:2159); real fading often does not follow a single exponential (Song et al. 1995, Biophys J 68:2588), and at high laser power many dyes bleach faster than in proportion to the power, so the total budget shrinks (Eggeling et al. 1998, Anal Chem 70:2651; Donnert et al. 2007, Nat Methods 4:81).

Remember

  1. 1Every exposure fades the dye a little and can harm living cells.
  2. 2Brighter or longer exposures buy a cleaner image, but spend the budget faster.
  3. 3Use the least light that answers your question.
Check yourself

You need 200 images of live cells over an hour. What should you change first?

Training

These pages are purely to improve your knowledge: practical training on the instrument is not only recommended but mandatory.