Learn · Guide · For when you write it up

Publishing your images

What goes in the Methods, what makes a figure trustworthy, and where the raw data should live.

Journals have started to ask. In a one-year pilot from June 2025, four Nature Portfolio journals asked authors invited to revise to complete a light-microscopy reporting table(opens in a new tab) designed with input from QUAREP-LiMi(opens in a new tab); Nature Communications still lists it among its author resources.⁠[1][6][13] This guide turns the community’s checklists⁠[2][3] into tools you can use with the Core’s instruments.

abcde MicrotubulesDNA Inset of a Microtubules DNA Microtubules, inset 5 µm5 µm5 µm1 µm1 µm 1430 photons per pixel
Figure 1 | A cell prepared to the checklist. a, Merge of microtubules (green) and DNA (magenta); the dashed box marks the area shown in b and e. b, The boxed area, enlarged by repeating pixels. c, d, Microtubules and DNA in grayscale. e, Microtubules in the boxed area, inferno lookup table, with a calibration bar in photons per pixel. Simulated confocal image (1 AU, NA 1.4 oil, 90 nm pixels, photon noise), not a micrograph. Display: linear, 0 to the brightest pixel of each channel (a–d) and 0 to 143 photons (e). Scale bars: 5 µm (a, c, d) and 1 µm (b, e).The legend itself is part of the example: it names every panel, channel, display range and scale. The numbers on the figure mark six things a reviewer looks for.⁠[3]
  1. A scale bar on every panel, with its length. A magnification does not tell the reader the pixel size.
  2. Channels named in the color they are shown in. Green and magenta stay distinct for color-blind readers.
  3. Each channel also in grayscale, where detail and intensity are easiest to judge.
  4. The inset’s position marked on the overview, and the inset enlarged by repeating pixels, not by interpolating them.
  5. A pseudocolor lookup table with a calibration bar in real units.
  6. Panels separated by white space, so no one reads two images as one.
1 · Why it matters

A reader should be able to repeat your image

Every setting in the lessons changes the picture: the objective’s NA sets the finest detail, the pinhole sets the section, the pixel size decides whether that detail is recorded, the detector and laser set the noise. A Methods section that leaves them out leaves the reader unable to judge the image, let alone repeat it.

QUAREP-LiMi(opens in a new tab), an international consortium of imaging scientists, core facility staff and manufacturers, has agreed what the minimum is. Its Working Group 11 lists what to report about the specimen, the hardware and the acquisition; Working Group 12 adds checklists for figures, image analysis and data. The Nature Portfolio reporting table(opens in a new tab) covers the same ground, plus the quality checks done on the instrument.⁠[1][2][3][6][13]

Why now
  1. 2004The Journal of Cell Biology sets out which image adjustments are acceptable and which are not.⁠[8]
  2. 2020A survey of published papers finds imaging methods widely underreported.⁠[5]
  3. 2021QUAREP-LiMi, the community consortium for quality in light microscopy founded in 2020, sets out its program.⁠[6]
  4. 2023Working Group 12 publishes checklists for image figures and image analysis (online September 2023; in print 2024).⁠[3]
  5. 2025Four Nature Portfolio journals pilot a light-microscopy reporting table.⁠[1]
  6. 2026Working Group 11 publishes the minimal requirements for reporting light microscopy.⁠[2]
01SpecimenCover glass (no. or thickness)Mounting or imaging mediumFluorophore or dye per channel02HardwareMicroscope stand and modalityObjective, as on the barrelLight source and wavelengthEmission selectionDetector type03AcquisitionPinhole (AU), dwell, averagingSequential or simultaneousFinal pixel sizez-step, time, tilingSoftware and version04Processing and analysisEvery step, in orderSoftware, version, settingsExample input and output05Figures and dataScale bars, channel namesSame display when comparedFiles shared openlyThe minimal Methods list (QUAREP-LiMi Working Group 11)Figure and analysis checklists (Working Group 12)01SpecimenCover glass (no. or thickness)Mounting or imaging mediumFluorophore or dye per channel02HardwareMicroscope stand and modalityObjective, as on the barrelLight source and wavelengthEmission selectionDetector type03AcquisitionPinhole (AU), dwell, averagingSequential or simultaneousFinal pixel sizez-step, time, tilingSoftware and version04Processing and analysisEvery step, in orderSoftware, version, settingsExample input and output05Figures and dataScale bars, channel namesSame display when comparedFiles shared openlyThe minimal Methods list (QUAREP-LiMi Working Group 11)Figure and analysis checklists (Working Group 12)
Diagram · What a reader needs, in the order a Methods section gives it, and in the order of this page: each column links to its section. The first three columns are the minimal list;⁠[2] the last two are the analysis and figure checklists.⁠[3] Metadata standards such as REMBI cover the same ground in machine-readable form.⁠[7]
2 · Write it

Your Methods paragraph

Pick the instrument and enter your settings. The paragraph follows the minimal checklist, in its order, and flags what is still missing.

01  Instrument

From the CoreInverted stand · four solid-state laser lines, for dyes excited at 405, 488, 561 and 640 nm · GaAsP detectors · Airyscan 32-element GaAsP array detector.

Detection

02  Objective

Immersion

03  Channels

2 of 4
Excitation line (nm)
Detection band
Excitation line (nm)
Detection band
Channels acquired

04  Acquisition

Averaged by
Pixels per line
Final pixel size81.9 µm ÷ 1024 px =0.080 µm
51 planes

05  Specimen

Cover glass

06  Processing

Processing
Generated · Methods textLight microscopy

Samples on #1.5H cover glass were mounted in [mounting medium]. Images were acquired on a Zeiss LSM 800 confocal laser-scanning microscope (inverted stand), using a 63×/1.40 oil-immersion Plan-Apochromat objective. EGFP was excited with the 488 nm solid-state laser and detected at 500–550 nm; mCherry was excited with the 561 nm solid-state laser and detected at 600–650 nm. Channels were recorded sequentially with GaAsP detectors, a 1.0 AU pinhole, a pixel dwell time of 2.0 µs and 2× line averaging; the final pixel size was 0.080 µm (1024 pixels across an 81.9 µm field). Z-stacks spanned 10 µm in 0.2 µm steps (51 planes). The acquisition software was ZEN. No deconvolution or denoising was applied.

What you enteredFrom the CoreStill to fill in
Minimal reporting items13 of 14
MissingMounting medium
Sampling checkComputed · textbook formulas
Pixel sizeMeets80 nmPractical rule: ≤ 89 nmHalf the lateral FWHM, 178 nm
Z-stepMeets200 nmPractical rule: ≤ 231 nmHalf the axial FWHM, 461 nm

Shortest excitation 488 nm · NA 1.40 · oil n 1.518. Lateral FWHM 0.51 λ/NA, axial 0.88 λ/(n − √(n² − NA²)); the practical rule takes half of each. Details in module 05 (pixels) and module 06 (z-steps).

1 minimal item missing: mounting medium. Still the example values: objective, channels, acquisition settings, z-stack. Pixel size 80 nm meets the practical rule (≤ 89 nm); the 200 nm z-step meets the axial rule (≤ 231 nm). Recommended: add the ZEN version (Help, About). Journals and the community checklists ask for it, and it lets others reproduce your settings.

The paragraph follows the items and order of the QUAREP-LiMi minimal reporting checklist(opens in a new tab)⁠[2]; journals may ask for more (for example the Nature Portfolio reporting table(opens in a new tab)⁠[13], or the modality-specific checklists of Montero Llopis et al.⁠[4]). Instrument descriptions come from the Core; the vendor’s acquisition software is filled in as a starting point, and everything else is what you enter, so check it against the file’s metadata. Sampling checks use the textbook formulas of modules 05 and 06 (pinhole ≥ 1 AU, one-photon excitation, shortest excitation wavelength). Final pixel size = scan field width ÷ pixels per line; planes = range ÷ step + 1; time points = duration ÷ interval + 1. Immersion indices as in the module calculators: oil 1.518, glycerol 1.47, water 1.333, air 1.0.

3 · Reporting the analysis

Every step between the file and the plot

Processing changes the numbers in an image, so it belongs in the Methods like any other step. What to write depends on where the workflow comes from: the newer it is to your readers, the more you share.⁠[3]

What to report, by kind of workflow

MinimalRecommendedIdealnot an item for that kind
What to report
Name it
MinimalMinimalMinimal
MinimalMinimalMinimal
Describe it
MinimalMinimalNot an item
Not an itemMinimalNot an item
RecommendedRecommendedNot an item
Not an itemRecommendedNot an item
Not an itemRecommendedRecommended
Share it
MinimalMinimalNot an item
MinimalMinimalMinimal
Not an itemNot an itemMinimal
Not an itemMinimalRecommended
RecommendedRecommendedRecommended
Not an itemNot an itemRecommended
Not an itemNot an itemMinimal
Not an itemNot an itemIdeal
Make it rerunnable
IdealIdealNot an item
IdealIdealRecommended
New workflow · existing components assembled into your own sequence

For a new workflow, the minimum is to cite every component and the platform with versions, describe the order of the steps, report the key settings, and share the workflow as code with example input, output and hand-drawn regions.

Recommended adds every setting, public data and code, the rationale and the limitations; ideal adds a tutorial and an easy install or container.

7 minimal
4 recommended
2 ideal
Row · hover or select

Cite the method and the platform

A citation (or the download link, author and date) lets a reader find exactly what you ran; the software platform is cited too.

Established workflow · Minimal
WG12: Cite workflow & platform
New workflow · Minimal
WG12: Cite components & platform
Machine learning · Minimal
WG12: Cite original method
Checklist · Levels as set by QUAREP-LiMi Working Group 12, in our words: minimal is non-negotiable, recommended is the goal to aim for, ideal is where the field is heading.⁠[3] Machine-learning levels apply to all models at the minimal level, to retrained and new models at the recommended level and to new models at the ideal level. Where the paper and the online book differ, each cell’s note says which one it follows.⁠[3][12] Choose a column for its minimum; hover over or select a row for the item behind each mark.
4 · Figures that survive review

A figure clinic

Nine rules from the community checklist for image figures, each with its level. Pick one and drag the divider: left of it, the figure as it is often submitted; right of it, as the checklist asks.⁠[3][12]

Rule 1 of 9 · Format

Inset: mark it, do not smooth it

Simulated

The checklist asks: Show exactly where an inset comes from in the full view, and enlarge it without an interpolation that suggests more resolution than was recorded.

Rule 1 of 9, Inset: mark it, do not smooth it (minimal). Left of the divider the figure as often submitted, right of it as the checklist asks. Inset position on the full view: unmarked versus dashed box. Inset screen pixels changed by smoothing: 84 % versus 0 %.
Left panel: the inset, 20× larger. Right panel: the full view. Drag the divider across both, or focus it and use the arrow keys.
Inset position on the full viewunmarked → dashed boxInset: 40 × 40 pixels, 3.6 µm wide, shown 20×
Inset screen pixels changed by smoothing84 % → 0 %Off by ½ photon or more from the pixel they cover

Smoothed and unplaced, the inset fills 84 % of its screen pixels with interpolated values that were never recorded and hides the 90 nm pixel grid, suggesting finer detail than was recorded; repeated pixels and a box on the full view show what was recorded, and where. A minimal requirement.

Simulated images: the cell used throughout the lessons, computed with a vectorial point-spread function for a confocal at 1 Airy unit, NA 1.4 oil, 488 nm excitation and 520 nm emission, 90 nm pixels, Poisson photon noise, no detector offset; microtubules (brightest pixel 143 photons) and DNA (88 photons). Rule 2’s field b is another cell of the same simulated specimen, scaled to the same peak brightness, with its own noise. Unless a rule is about the display range, each channel is shown linearly from 0 to its brightest pixel. 1: the inset is 40 × 40 pixels shown 20× larger; “changed by smoothing” counts screen pixels where bilinear interpolation (pixel-centered) differs by half a photon or more from the pixel underneath. 5: relative luminance of the sRGB primaries (green 0.7152; magenta = red + blue, 0.2126 + 0.0722). 3: the narrow range ends at the 99th percentile of the image (the smallest value that at least 99 % of pixels do not exceed); “clipped” means brighter than the display maximum. 4: deuteranopia simulated with the Machado, Oliveira and Fernandes model at full severity on linear RGB;⁠[10] ΔE is CIE76 in CIELAB (D65), for each color at full brightness against the other at its most similar brightness between 20 and 100 %. 6: matplotlib’s inferno lookup table. 7–8: print sizes assume the panel is printed 40 mm wide (1 pt = 0.353 mm); contrast is the WCAG ratio of relative luminances, (L1 + 0.05) / (L2 + 0.05), against black. 9: properties are typical. Vendor files, OME-TIFF and TIFF can also be saved with lossy compression (for example JPEG XR in .czi); check the save options. The JPEG evidence is a real round trip: a 64 × 64 crop exported as 8-bit green RGB (0 to 143 photons → 0 to 255), saved with Pillow (libjpeg) at quality 75 with 4:2:0 chroma subsampling, decoded and converted back to photons; the PNG size is the same 8-bit export saved losslessly. Levels as given in the QUAREP-LiMi Working Group 12 image-publishing checklists, online book (2024 version): all minimal here except the calibration bar, recommended; rules paraphrased from them. The 2024 paper sets a lower minimum for data: authors prepared to share the images shown or measured.⁠[3][12]

Measure it yourself

Three ways a figure can mislead

Each demo runs on the simulated cell from the lessons. Change one setting and compare what a reader would see with what the numbers say.

Demo A · simulated

Same display, or auto-contrast?

A control cell and a “treated” cell that gives 35 % of the control’s photons. Images that are compared must be shown with the same display settings.⁠[3]

Top: the two cells as displayed. Below each: its histogram (photons per pixel, log scale) and the lookup table that maps it to the screen.
Display
Mean in the cell, control10.5 photonsShown as 0–94 photons
Mean in the cell, treated3.7 photonsShown as 0–34 photons, its own range
Photons, treated ÷ control0.35×What the detector counted
On screen, treated ÷ control0.97×Misleading: the screen exaggerates the treated cell

Auto-contrast stretched each image to its own range (0–94 and 0–34 photons), so on screen the treated cell looks 0.97× as bright as the control, although it gave only 0.35× the photons; only its grainier noise gives it away. Choose “Same display range”.

Demo B · simulated

Will a color-blind reader see it?

The same two channels merged in three color pairs, seen with typical color vision or simulated as seen with one of the three dichromacies.⁠[10] In a fluorescence image brightness varies everywhere, so channels must differ in hue, not only in brightness.

Left: the merge (microtubules and DNA) as this reader sees it. Right: each color from dim to bright, and the closest match between them.
Colors (microtubules / DNA)
Seen with
Microtubules vs DNAΔE 6.7closest match (28 with both at full brightness)
Overlap vs microtubulesΔE 4.0closest match, yellow vs green
Overlap vs DNAΔE 3.5closest match, yellow vs red

Seen with deuteranopia, green and red differ mainly in brightness: the closest match is ΔE 6.7, about 3 just-noticeable differences, too little for thin, noisy structures. Use green and magenta.

Demo C · simulated

Rotate or compress, then measure

A 5.8 µm patch of microtubules near the nucleus, in photon counts. Rotating by an angle that is not a multiple of 90° interpolates new pixel values; JPEG throws detail away to make the file small. Both change what you would measure.⁠[3][9]

Left: the original. Middle: rotated by 15°. Right: rotated back minus the original (orange: values gained, blue: values lost). Statistics inside the dashed circle.Bottom: every pixel before against after (a perfect copy falls on the dashed line), and the values along the orange row.
Operation
Rotation angle15°
Rotating by 90°, 180° or 270° moves pixels without changing them.
Pixel values changed79 %after rotating back by 15°
Largest change−16 photonsin one pixel, of 67 at the brightest
Brightest pixel67 → 64−4 %, in the rotated image
Total intensity+0.1 %inside the dashed circle

Rotating by 15° and back changes 79 % of the pixel values (by up to 16 photons), and the rotation alone lowers the brightest pixel from 67 to 64 photons (−4 %). Measure before you rotate, or rotate the finished panel in vector software.

Simulated images: the cell used throughout the lessons, computed with a vectorial point-spread function for a confocal at 1 Airy unit, NA 1.4 oil, 488 nm excitation and 520 nm emission, sampled at 90 nm pixels with Poisson photon noise and no detector offset. A: the treated image is the same cell at 0.35× the expected photons with its own noise; “the cell” is the area where the noise-free control gives at least 2 photons; auto-contrast here maps each image’s minimum to black and its 99.9th percentile to full brightness; “on screen” compares mean displayed values (0 to 1), a linear measure, not perceived brightness. B: Fiji’s standard lookup tables (pure green, red, magenta, cyan), merged by adding the 8-bit color values; dichromacy simulated with the Machado, Oliveira and Fernandes model at full severity,⁠[10] applied to linear RGB (sRGB decoded, matrix, clipped, encoded); color difference is CIE76 ΔE*ab in CIELAB (D65 white, sRGB primaries). The closest match compares each color at full brightness with the other at 20 to 100 % of its display range. A difference of about 2.3 is often quoted as just noticeable for large uniform patches;⁠[11] thin, noisy structures are harder to judge, so below 10 (about four just-noticeable differences) we call a pair easily confused. Typical vision shows the colors unchanged. C: bilinear interpolation about the patch center (zero outside the patch), results rounded to whole numbers as an integer image stores them; statistics within the dashed circle (26 pixels radius), where every rotated value comes from inside the patch; changes are measured after rotating back by the same angle. JPEG: the patch exported as an 8-bit RGB file in the green lookup table (0 to its brightest pixel → 0 to 255), saved with Pillow (libjpeg) at the stated quality with 4:2:0 chroma subsampling, decoded, and the green channel converted back to photons; the PNG size is for the same file saved losslessly. The rules come from the QUAREP-LiMi image-publishing checklists(opens in a new tab).⁠[3][12]

Image integrity

Honest adjustments

The rule behind every journal policy: the published image must faithfully represent the data. Most problems are not fraud but well-meant adjustments that went too far.⁠[8][9]

Generally acceptable

  • Adjusting brightness and contrast linearly, over the whole image.
  • Applying the same adjustment to every image in a comparison, and to its controls.
  • Cropping, and rotating by multiples of 90°.
  • Non-linear adjustments such as gamma, when they are disclosed in the figure legend.
  • Deconvolution or denoising applied identically to every compared image, checked not to create or remove structures, and stated in the Methods (and in the legend where it matters).

Not acceptable

  • Changing one part of an image differently from the rest: erasing, cloning, brightening a region.
  • Combining images from different fields, samples or times without a visible boundary and a statement.
  • Adjusting so that faint signal disappears or background becomes signal.
  • Choosing a non-representative image, or hiding the processing that made the result.
  • Auto-contrasting each image of a comparison separately, so that different signals look the same.

Is this adjustment acceptable?

Simulated

Left, what a reader sees. Right, a forensic noise map. Photon noise has a fingerprint: the variance of the counts equals their mean. An adjustment applied to the whole image changes that ratio by the same rule everywhere (for gamma, smoothly with brightness); an adjustment applied to one region leaves its outline in the map.⁠[8][9]

Dashed: the outline of the edited area (for the whole-image adjustments, the same area for comparison). Grey in the map: signal, where noise cannot be measured.
Adjustment
Noise ratio inside2.03variance ÷ mean, background
Noise ratio outside1.02variance ÷ mean, background
Inside ÷ outside2.00×a region stands out
NOT ACCEPTABLE

Not acceptable: one part of the image was changed differently from the rest. Brightening ×2 made the variance 4× and the mean 2× inside the outline only, so the ratio is 2.03 there against 1.02 outside: a sharp-edged region in the map, the kind of discontinuity forensic image checks look for.

Simulated: the cell field used in the lessons (vectorial point-spread function, confocal at 1 Airy unit, NA 1.4 oil, 488/520 nm), 30 nm pixels averaged 3 × 3 to 90 nm, scaled so that the brightest 0.1 % of pixels reach about 60 photons, plus a uniform background of 1.5 photons, with one fixed draw of Poisson noise. The adjustments act on the values in floating point (no 8-bit rounding): linear, every pixel × 2; gamma, v = 120 (n/120)0.6; one cell, × 2 inside the dashed outline only. Every image is displayed from 0 to 120. Noise map: for each pixel, the variance (÷ N) of the 15 × 15 pixels around it divided by their mean, shown relative to the median over the background (neutral from ×0.84 to ×1.19, the range of the estimate’s own scatter). Windows with signal (local mean above 2.5 × the image’s median local mean) and pixels within 7 pixels of them are not measured (grey). “Inside” and “outside”: mean variance ÷ mean of the means, over measured pixels whose window lies entirely inside, or entirely outside, the outline. For Poisson counts the ratio is 1; multiplying by k makes it k. Real images add detector gain and offset, read noise and compression, so the reference value differs, but a region edited differently from the rest still stands out. Numbers reproduced in numpy by pub_extras_check.py. Rules: image-integrity guidance⁠[8][9] and the WG12 checklists.⁠[3]

5 · Keep the originals, share the data

The figure is not the data

The file the microscope writes (.nd2, .czi, .lif) holds the pixel values and the settings that made them. Keep it untouched and work on copies. Rule 9 of the clinic shows what each format keeps and what a single JPEG save does to the numbers.⁠[3]

“Available on request” has been shown not to work well, and the checklists discourage it. Making the images shown and measured available is the minimum; a public repository with an open license is the recommended level; the Nature Portfolio journals point authors to the BioImage Archive, the Image Data Resource, BioStudies and figshare.⁠[1][3]

6 · Before you submit

Check it before the reviewer does

Tick what your manuscript already does. The levels follow Working Group 12: minimal is the requirement, recommended the goal, ideal where the field is heading.⁠[2][3]

0 of 17
Nothing ticked yet. Start with the Methods items: the builder in section 2 writes most of them for you.

Methods

0 of 8 · WG11

Figures

0 of 6 · WG12

Analysis

0 of 2 · WG12

Data

0 of 1 · WG12

“Copy submission pack” copies your Methods paragraph from section 2, the items still open at this level, and the acknowledgment sentence, ready to paste into your manuscript notes.

7 · Credit the Core

One sentence in the acknowledgments

Acknowledgments are how a core facility shows its funders what its instruments made possible; that record keeps them funded and running. Please include this sentence in every paper, poster or thesis that uses data from the Core.

When Core staff did more than help with your imaging, for example designing the experiment, acquiring the images, analyzing the data or writing the paper, talk to us about co-authorship.

Acknowledgment

Imaging was performed at the UNLV Confocal and Biological Imaging Core.

Used the Nikon A1R MP? Add: “The multiphoton system was acquired with support from the National Science Foundation, award DBI-1726925.”

References

The checklists are summarized here in our own words. Read the originals before you submit: they are short, open access and updated by the community.

  1. [1]Light microscopy reporting for reproducibility. Editorial. Nature Cell Biology 27:877, 2025. Announces the cross-journal light-microscopy reporting table (Nature Communications, Nature Cell Biology, Nature Methods, Nature Structural & Molecular Biology). doi:10.1038/s41556-025-01704-y
  2. [2]Montero Llopis P, van Oostende-Triplet C, et al. Better reporting is better science: community-defined minimal reporting requirements for light microscopy. Journal of Cell Biology 225(3):e202601032, 2026. QUAREP-LiMi Working Group 11. doi:10.1083/jcb.202601032
  3. [3]Schmied C, Nelson MS, Avilov S, et al. Community-developed checklists for publishing images and image analyses. Nature Methods 21:170–181, 2024. QUAREP-LiMi Working Group 12. doi:10.1038/s41592-023-01987-9
  4. [4]Montero Llopis P, Senft RA, Ross-Elliott TJ, et al. Best practices and tools for reporting reproducible fluorescence microscopy methods. Nature Methods 18:1463–1476, 2021. doi:10.1038/s41592-021-01156-w
  5. [5]Marqués G, Pengo T, Sanders MA. Imaging methods are vastly underreported in biomedical research. eLife 9:e55133, 2020. doi:10.7554/eLife.55133
  6. [6]Boehm U, Nelson G, Brown CM, et al. QUAREP-LiMi: a community endeavor to advance quality assessment and reproducibility in light microscopy. Nature Methods 18:1423–1426, 2021. doi:10.1038/s41592-021-01162-y
  7. [7]Sarkans U, et al. REMBI: Recommended Metadata for Biological Images—enabling reuse of microscopy data in biology. Nature Methods 18:1418–1422, 2021. doi:10.1038/s41592-021-01166-8
  8. [8]Rossner M, Yamada KM. What’s in a picture? The temptation of image manipulation. Journal of Cell Biology 166:11–15, 2004. doi:10.1083/jcb.200406019
  9. [9]Cromey DW. Avoiding twisted pixels: ethical guidelines for the appropriate use and manipulation of scientific digital images. Science and Engineering Ethics 16:639–667, 2010. doi:10.1007/s11948-010-9201-y
  10. [10]Machado GM, Oliveira MM, Fernandes LAF. A physiologically-based model for simulation of color vision deficiency. IEEE Transactions on Visualization and Computer Graphics 15:1291–1298, 2009. doi:10.1109/TVCG.2009.113
  11. [11]Mahy M, Van Eycken L, Oosterlinck A. Evaluation of uniform color spaces developed after the adoption of CIELAB and CIELUV. Color Research and Application 19:105–121, 1994. doi:10.1111/j.1520-6378.1994.tb00070.x
  12. [12]QUAREP-LiMi Working Group 12. Checklists for image publishing, with a Fiji course “How to publish images”. Online book, quarep-limi.github.io/WG12_checklists_for_image_publishing.
  13. [13]Springer Nature. Light microscopy reporting table (spreadsheet), linked from the Nature Communications author resources: nature.com/documents/Light_microscopy_reporting_table.xlsx.