The instrument, assembled

A generic inverted laser-scanning confocal on an optical table: a white stand with binocular eyepieces, a dark transmitted-light pillar above the stage and the objectives below it, the white scan head module docked to the stand's left side port, and around it the laser and control units, a joystick and a computer.
A generic inverted laser-scanning confocal. Schematic model, not to scale.

Follow one photon

Press Play or space to watch the film, and ← → to jump between parts. Prefer to go at your own pace? Switch to Scroll, and the film advances as you scroll.

0% loaded

Laser — single-wavelength light.

00.0 s · 1 / 11

The eleven parts

Scroll and the instrument comes apart; the light path is then drawn through the parts. Select a number, on the picture or in the list, to read about that part.

00.0 s · assembled

Part by part

What each part does, why it matters, and what goes wrong when it is set badly.

1 / 11

Laser

What it does
Supplies light at the wavelength that excites the fluorophore, in a narrow, intense beam. Most systems combine several laser lines, each for a different group of dyes.
Why it matters
The excitation wavelength decides which dyes you can image. Power is adjustable, and every bit of extra power adds photobleaching.
When it goes wrong
Too much power bleaches and damages the sample; too little gives a dim, noisy image.

2 / 11

Dichroic mirror

What it does
Reflects the excitation light toward the scanner and lets the longer-wavelength fluorescence pass through toward the detector.
Why it matters
It separates the faint fluorescence from the far brighter laser light that travels the same path.
When it goes wrong
A dichroic chosen for other dyes reflects the wrong wavelengths: you lose signal, or excitation light leaks into the detector.

3 / 11

Scan mirrors

What it does
Two galvanometer mirrors, one for X and one for Y, steer the focused spot across the sample line by line. The returning fluorescence reflects off the same mirrors on its way back, which is called descanning.
Why it matters
Scan speed sets how long the spot stays on each pixel, and that sets how many photons each pixel collects. Descanning keeps the emitted light still at the pinhole while the spot moves.
When it goes wrong
Scanning too fast leaves few photons per pixel and a noisy image. Scanning too slowly adds bleaching and drift.

4 / 11

Scan lens

What it does
Works with the tube lens to relay the scanner into the microscope, so the beam pivots about the back aperture of the objective.
Why it matters
Because the beam pivots at the back aperture, it stays centered on the objective at every scan angle, so the field is evenly bright and sharp out to its edges.
When it goes wrong
A poor match between scan lens and tube lens vignettes the beam, and the edges of the field go dark.

5 / 11

Tube lens

What it does
Works with the scan lens. The two lenses form an intermediate image between them and relay the scan mirrors onto the back aperture of the objective. The returning fluorescence retraces the same path.
Why it matters
With the objective, it sets the magnification. The pinhole is not at its image plane: after descanning, a separate pinhole lens focuses the light onto the pinhole, in a plane conjugate to the focus.
When it goes wrong
A tube lens that does not match the objective gives the wrong magnification and adds aberrations, so the image is less sharp and dimmer.

6 / 11

Objective

What it does
Focuses the excitation light into a small spot, and collects the emitted light back.
Why it matters
Its numerical aperture sets both the resolution and how much light it collects. Higher NA is brighter and sharper.
When it goes wrong
A wrong immersion medium, or a sample deeper than the design allows, causes spherical aberration: the image blurs and dims.

7 / 11

Sample and focal plane

What it does
The fluorescent specimen. The laser excites fluorophores above and below the focus too; the pinhole rejects most of their light, so each image is an optical section.
Why it matters
Moving the focus through the sample builds a stack of optical sections, the raw material for every 3D view.
When it goes wrong
Focus drift, or imaging deep in a thick, scattering sample, gives blurred or dim planes.

8 / 11

Pinhole

What it does
A small aperture in a plane conjugate to the focus. Light from the focal plane passes; most light from above and below is blocked.
Why it matters
This is what makes a confocal image a confocal image. Its size is set in Airy units; about 1 AU is the usual compromise.
When it goes wrong
Opened wide, out-of-focus blur returns. Closed too far, the image becomes dim and noisy.

9 / 11

Filter or spectral selector

What it does
Chooses which wavelengths reach the detector. Some systems use a prism or grating with an adjustable slit instead of a fixed filter.
Why it matters
A band matched to one dye's emission helps keep each dye in its own channel.
When it goes wrong
A band that is too wide lets one dye's light into another channel. This is bleed-through.

10 / 11

Detector

What it does
Turns the few photons that arrive into an amplified electrical signal, read out once per pixel.
Why it matters
PMT, GaAsP and hybrid detectors differ in sensitivity and noise. This decides how faint a signal you can see.
When it goes wrong
High gain cannot replace missing photons: it amplifies the noise along with the signal. Pixels that saturate lose the true intensity.

11 / 11

Computer

What it does
Records one intensity value per pixel as the spot scans, and assembles the image.
Why it matters
Stacking optical sections gives a Z-stack; a 3D view is rendered from that stack.
When it goes wrong
Pixels that are too large undersample fine detail (aliasing). Pixels that are too small waste time and light.

The captions, in order

  1. 0–3 s Laser — single-wavelength light.
  2. 3–6 s Dichroic mirror — reflects excitation, transmits fluorescence.
  3. 6–10 s Scan mirrors — steer the spot across the sample, line by line.
  4. 10–13 s Scan lens and tube lens — relay the scanner into the objective.
  5. 13–17 s Objective — focuses the laser and collects the fluorescence.
  6. 17–22 s Sample — fluorescence from the focus, and from above and below it.
  7. 22–26 s Descanning — the returning light is held still by the same mirrors.
  8. 26–29 s Dichroic mirror — the fluorescence passes through.
  9. 29–33 s Pinhole — in-focus light passes, most out-of-focus light is blocked.
  10. 33–36 s Emission filter — only the fluorescence band continues.
  11. 36–38.5 s Detector — photons become an electrical signal.
  12. 38.5–42 s Computer — one value per pixel builds the image.