Module 09 of 10 All modules
Module 09 · Chapter V · Beyond the single pinhole

Multiphoton excitation

Why squared intensity confines excitation to the focus, and what that buys in depth and gentleness, at a cost in resolution.

live model
Computed one-photon excitation at 488 nm, x–z section through the focus: the light forms a double cone above and below the focus1-photon, 488 nm
Computed two-photon excitation at 976 nm, x–z section through the focus: excitation is confined to a small spot at the focus2-photon, 976 nm
COMPUTED · Where excitation happens, x–z section, 4.8 × 6 µm, NA 1.0 water: intensity for one photon, intensity squared for two, each to its own peak. Brightness on a power-law scale (γ 0.35) so the cones show.
1 · The physics

A pinhole made of physics

A fluorophore that absorbs blue light can also be excited by two near-infrared photons arriving almost simultaneously, each carrying about half the energy. The probability scales with the square of the intensity, so excitation happens only where photons are extremely dense: at the focus of a high-NA objective illuminated by a femtosecond pulsed laser, which packs the photons into brief, intense bursts at a low average power.[14][15]

Away from the focus the intensity falls and its square falls faster: integrated over a plane, one-photon excitation is the same at every distance from the focus, two-photon excitation falls as 1/z² and is down to 7 % two micrometres away. The excitation volume is confined in three dimensions by the illumination itself, and no pinhole is needed. Every emitted photon is useful, including scattered ones, so the detector collects all the light it can, without a pinhole (non-descanned detection). This is what makes depth possible: in scattering tissue, high-resolution one-photon imaging is limited to roughly the first 100 µm, while two-photon imaging reaches several hundred micrometres, up to about 1 mm in neocortex, because near-infrared light is scattered less and the scattered emission still counts.[15] Above and below the focus there is almost no excitation, which spares those planes from bleaching and photodamage; inside the focal volume, though, bleaching and photodamage can rise more steeply than the square of the intensity.[15]

The cost is resolution: the excitation wavelength is about twice as long, and squaring the intensity only partly compensates. In our model at NA 1.0 in water, the two-photon focus is 378 nm wide and 1388 nm long, against 264 nm and 964 nm for one-photon excitation at 488 nm.[20] These one-photon numbers are for the excitation spot alone, which is close to what a confocal delivers with its pinhole at about 1 AU (module 03). The textbook √2 advantage of an ideal confocal needs a nearly closed pinhole, which is why in practice two-photon and confocal resolution are closer than ideal theory suggests.[15] The same lasers drive second-harmonic generation in ordered, non-centrosymmetric structures such as collagen, giving label-free signal at exactly half the excitation wavelength.[15][18]

COMPUTED · ENERGIES TO SCALE (E = hc/λ) · T₁ NOT TO SCALE012energy (eV)S₀S₁≈ 2.48 eVestimated 0–0vibrational levelsT₁not to scaleintersystem crossingrare: competes withemission, within nsµs or longerlong-lived, so it has timeto react, e.g. with oxygen:for many dyes a main routeto bleaching (module 08)one photon488 nm2.54 eV~fs~pstwo photons976 nm1.27 eV976 nm1.27 eVvirtual statenot a real level: the two photonsmust arrive practically togetherrate ∝ I²emission507 nm (peak)2.45 eVa band: 497–528 nmat half heightlifetime 2.6 ns
COMPUTED · Energies to scale, E = hc/λ. S₁ is drawn at EGFP’s 0–0 energy, estimated as 2.48 eV (501 nm) from where its excitation and emission spectra cross (FPbase 1 nm data); a 488 nm photon lands 0.064 eV above it and the molecule relaxes in picoseconds; emission at the 507 nm peak (FPbase’s listed peak; the measured curve peaks at 511 nm) lands 0.031 eV above S₀, which relaxes too. Module 01 draws the same loss, 0.10 eV peak to peak, as a single step on the S₁ side. Two 976 nm photons (2 × 488 nm, the wavelength module 09’s model uses) reach the same state through a virtual state, with a probability that grows as the intensity squared; EGFP’s measured two-photon spectrum actually peaks at a shorter wavelength (Drobizhev et al. 2011, Nat Methods 8:393). Time scales are orders of magnitude (Lakowicz 2006, Principles of Fluorescence Spectroscopy, ch. 1), except EGFP’s lifetime (2.6 ns, FPbase); the triplet level is not to scale, and its lifetime, a microsecond or longer, varies with the dye and its surroundings (Donnert, Eggeling & Hell 2007, Nat Methods 4:81). The vibrational bands are drawn schematically.
2 · Live model

How deep can each mode see?

Tissue scatters light, and scattered light misses the focus. Move the focus deeper into brain gray matter and compare one- and two-photon excitation: where fluorescence is made, what the detector can still use, and how much laser power it takes.

Focus depth below the tissue surface300 µm
Medium
1-PHOTON 488 nm2-PHOTON 976 nm
Excitation reaching the focus unscattered0.25 %22 %
Focal fluorescence, same surface power0.25 %5.0 %
Share of it the detector can use0.25 %all (assumed)
Detected focal signal, same power6.1·10⁻⁴ %5.0 %
Surface power for the same detected signal× 1.6·10⁵× 4.5
Made outside the ±2 µm focal slab ÷ inside5,041×0.16×

At 300 µm, 0.25 % of the 488 nm light reaches the focus unscattered, 22 % of the 976 nm light. For the same detected signal a confocal needs 1.6·10⁵ times the surface power, a two-photon microscope 4.5 times. The first is far out of reach; the second is routine.

Surface laser power needed to keep the detected focal signal (log scale) < 100 µm≈ 1 W: 600–800 µm020040060080010001200110²10⁴10⁶focus depth (µm)surface power, relative to focusing at the surfaceone-photon confocal: light lost going in and coming outtwo-photon: light lost going in only
Fluorescence made outside the focal slab, relative to the slab (uniformly stained tissue 1.5 mm thick; log scale) 0200400600800100012000.111010²10³10⁴focus depth (µm)made outside the focal slab ÷ insideone-photon: the whole cone glowstwo-photon: 1/z² wings, then near-surface lightas much outside as inside

Model[15]: only unscattered (ballistic) light reaches the focus, Pball = P0 e−z/ℓs, so the two-photon focal signal falls as e−2z/ℓs. Scattering mean free paths, brain gray matter: measured 50–100 µm at 630 nm in extracted tissue and about 200 µm at 800 nm in vivo[15]. We use ℓs = 50 µm for 488 and 520 nm (scattering is stronger at shorter wavelengths, so if anything this flatters one-photon imaging) and 200 µm for 976 nm (longer wavelengths scatter less, so this is conservative for two-photon imaging). NA 1.0 water immersion (cone half-angle 48.6°); every ray is attenuated over the depth z (oblique rays travel further and are lost sooner, which narrows the effective NA with depth; ignored). No absorption, aberration or refractive-index mismatch; uniformly stained tissue 1.5 mm thick (the maps show the top 1.2 mm). Fluorescence made per plane: one-photon ∝ ballistic power; two-photon ∝ plane integral of I², from the vectorial focus[20] out to 6 µm and (0.237 + 0.094/z)/z² beyond (fitted to the same calculation out to 96 µm). Focal slab ±2 µm. Scattered excitation light is ignored; for one-photon excitation it also excites dye, so the out-of-focus totals are lower bounds. Detection: the confocal pinhole passes only emission that leaves the focus unscattered (e−z/ℓs at 520 nm); the non-descanned detector is assumed to collect scattered emission as well as unscattered (no collection loss with depth). Each mode is shown relative to its own value at the focus in a clear medium; absolute one- versus two-photon brightness is not compared. The x–z maps colour each plane of the cone by the fluorescence made in that whole plane (log scale over eight decades), not by brightness per unit area; insets: computed focal excitation, own peak, linear. Shaded in the power chart: high-resolution one-photon imaging is limited to less than about 100 µm in tissue, and 1 W lasers reach about 600–800 µm in neocortex[15].

Formulas used in this module

The math, with its sources

Two-photon excitation rate[14][18]
R2P ∝ σ2 ⟨I²⟩
σ2: two-photon cross-section. Quadratic in intensity, so confined to the focus.
Why pulses help[15][18]
⟨I²⟩ ≈ ⟨I⟩² / (f τ)
Pulses of duration τ at repetition rate f (up to a pulse-shape factor of order one): 100 fs pulses at about 100 MHz give a two-photon advantage of about 10⁵ over continuous light of the same average power.
Plane-integrated excitation[14][20]
∬ I dx dy = const;∬ I² dx dy ∝ 1 / z²
One-photon excitation is the same in every plane; two-photon excitation falls with the beam area. Computed for NA 1.0 at 976 nm: 7 % of the focal-plane value at 2 µm, ≈ 0.237 µm² / z² far from focus.
Ballistic light with depth[15]
Pball = P0 e−z/ℓs;F2P ∝ e−2z/ℓs
Only unscattered photons focus. ℓs: scattering mean free path in brain gray matter, about 200 µm at 800 nm in vivo and 50–100 µm at 630 nm in extracted tissue. Holding the focal signal needs P0 ∝ ez/ℓs.
Confocal detection at depth[15]
S1P ∝ P0 e−z/ℓs(λex) e−z/ℓs(λem)
The pinhole accepts only emission that appears to come from the focus; beyond about one scattering length, unscattered emission is a minor part of what leaves the tissue. Our model keeps only the unscattered part.

About the model

The focal numbers are computed for a transparent, index-matched medium with the vectorial model.[20] The depth model adds scattering in the simplest form: only unscattered light reaches the focus, with scattering lengths measured in brain gray matter; scattered excitation light, absorption, aberrations and the narrowing of the effective NA with depth are left out; the pinhole passes only unscattered emission and the non-descanned detector collects all emission; the tissue is uniformly stained. Real tissue is inhomogeneous, and the depth limit also depends on laser power, pulse width, collection optics and the staining pattern.[15]

At the microscope

Multiphoton is the tool for depth and for gentle long-term imaging of thick living tissue, not for the finest detail in a thin section.

Plan on raising the laser power with depth, about ez/ℓs; the power modulator can ramp it automatically with focus position.[15]

Many dyes and fluorescent proteins have two-photon absorption peaks that are not simply twice their one-photon peaks; check measured two-photon spectra.[18]

Check yourself

Why doesn’t a two-photon microscope need a pinhole for optical sectioning?

References

Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.

  1. [14]Denk W, Strickler JH, Webb WW. Two-photon laser scanning fluorescence microscopy. Science 248:73–76, 1990. doi:10.1126/science.2321027
  2. [15]Helmchen F, Denk W. Deep tissue two-photon microscopy. Nature Methods 2:932–940, 2005. doi:10.1038/nmeth818
  3. [18]Zipfel WR, Williams RM, Webb WW. Nonlinear magic: multiphoton microscopy in the biosciences. Nature Biotechnology 21:1369–1377, 2003. doi:10.1038/nbt899
  4. [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.