Enhanced resolution: Airyscan and LIGHTNING
Image scanning microscopy with pixel reassignment, and adaptive deconvolution: what each gains, computed with our model, and the conditions each needs.
Small pinholes, without throwing light away
Module 03 showed the dilemma of the single pinhole: closing it toward 0.2 AU sharpens the image but discards most of the light. Image scanning microscopy replaces the pinhole with an array of small detector elements. Each element behaves as a small, slightly off-axis pinhole: it sees a sharp image, displaced from the scan position by about half its own offset. Shifting each element’s image back before summing, pixel reassignment, keeps the small-pinhole sharpness while collecting as much light as a 1 AU pinhole, or a little more (1.11× in our model, since the array covers about 1.2 AU). The idea was proposed by Sheppard in 1988 and demonstrated by Müller and Enderlein in 2010.[16]
Our model reproduces the gain: with 31 elements of 0.2 AU, reassignment alone gives a lateral FWHM of 141 nm against 187 nm at 1 AU (about 1.32× narrower), while collecting 1.11× the 1 AU signal. A single 0.2 AU pinhole reaches 147 nm but collects only 8 % of the 1 AU signal from a scanned point: its point image peaks at 0.13 of the 1 AU peak, the reassigned array’s at 1.77.[20]
Airyscan is Zeiss’s implementation, with 32 GaAsP elements covering about 1.25 AU, followed by a deconvolution step. Zeiss states 1.7× higher resolution in x, y and z than a conventional confocal; the part beyond pixel reassignment comes from the processing.[17]
One pinhole, a small pinhole, or an array
The point image of a 1 AU pinhole, a 0.2 AU pinhole and a 31-element array after pixel reassignment, computed with the same vectorial model and drawn on one brightness scale, with the light each collects underneath, so that sharpness and light can be judged together.
Light collected is the area under each point image (its plane integral): the bars under the pictures. Brightness is light per pixel.
On one scale the trade-off is plain: the 0.2 AU pinhole is sharp (147 nm) but collects 8 % of the light; the array keeps that sharpness (141 nm) and collects 1.11× the light of a 1 AU pinhole, because every element is shifted into place before the sum.
Each element sees a sharp, displaced spot
All 31 elements: summed as recorded they act as one large pinhole about 1.2 AU wide (192 nm); shifted back by α·d first they give 141 nm with 1.11× the 1 AU light. That is the whole gain of pixel reassignment, before any deconvolution.
Vectorial point spread functions[20], NA 1.4 oil, 488 nm excitation, 520 nm emission, no noise, no deconvolution. Array: 31 hexagonal elements of 0.2 AU (90.6 nm in the sample) on a hexagonal grid within 0.55 AU of the axis; each element’s image is shifted by α·d toward the axis, α = λexc²/(λexc² + λem²) = 0.47, then summed. Point images are shown as computed, on a linear scale: in “common scale” every image is relative to the same peak (the reassigned array, 1.77× the 1 AU peak), so brightness is proportional to light per pixel and the bars, drawn to one scale, to the light collected (plane integral). The central element alone collects 0.091 of the 1 AU signal, slightly more than a round 0.2 AU pinhole (0.083), because a hexagon of the same width has 10 % more area. Point images are drawn from 10 nm samples; the numbers come from the full computation on a 4 nm grid.
Sharper by computation, within limits
LIGHTNING takes a different route: an ordinary confocal detector and pinhole, with pixels fine enough for the detail to be recovered, followed by deconvolution. The recorded image is modeled as the true object blurred by the point spread function plus noise, and an iterative algorithm estimates the object that best explains the data. Leica describes its algorithm as adaptive: it assesses signal and noise locally and chooses parameters region by region, which protects dim regions from being sharpened into noise. Leica states lateral resolution “down to 120 nm”.
Deconvolution of any kind has conditions: Nyquist sampling in all three dimensions, a point spread function that matches the objective, wavelength and pinhole used, and enough photons. The live model below runs the classic Richardson–Lucy algorithm on a two-dimensional test object blurred by our computed 1 AU confocal point spread function, at three photon levels and with a matched or a mismatched point spread function, so that the gain and the two classic failure modes, amplified noise and a wrong blur model, can be seen.[24][10]
Neither Airyscan nor LIGHTNING reaches the resolution of STED or single-molecule localization; both are best described as enhanced-resolution confocal, with a strong gain in signal-to-noise for Airyscan. For separations of roughly 150 to 250 nm, below the confocal limit but well above what STED or localization microscopy are needed for, that is often exactly what is needed.[19]
- Fine enough sampling in x, y and z, at or near Nyquist (module 05, 06).
- A matching point spread function: same objective, wavelength, pinhole and refractive index, measured with beads or computed (module 04).
- Enough photons: noise, not the algorithm, sets how far the image can be sharpened (module 08).
What it cannot do: it can only restore contrast the optics transmitted strongly enough to stand above the noise, and nothing beyond the optical cutoff.
Between 182 nm and the cutoff a smaller pinhole or pixel reassignment (§1) lifts the curve; beyond 89.9 nm nothing passes with any pinhole. Module 05 shows the contrast of a bar target through the same optics, at your pixel size.
Deconvolution: the gain and the failure modes
A test object blurred by the computed 1 AU confocal point spread function, with shot noise. Run Richardson–Lucy iterations on it: watch two filaments 200 nm apart separate, then see what too few photons, or a point spread function that does not match the microscope, make of the same iterations.
With plenty of photons and the exact PSF the iterations separate the pair (dip 52 %, peaks 208 nm apart) and beading stays low (2 %).
Object: 128 × 128 pixels of 40 nm (5.1 µm field): 25 nm-wide filaments (two of them 200 nm apart), 40 nm points in pairs 160, 200 and 280 nm apart, drawn as a test target, not a cell. Blur: the computed vectorial 1 AU confocal point spread function[20] (NA 1.4 oil, 488/520 nm, lateral FWHM 187 nm) integrated over the same 40 nm pixels, 33 × 33 pixels, circular convolution. Noise: Poisson, scaled so that the brightest noise-free pixel (where two filaments cross) receives the chosen mean number of photons; single filaments peak at 0.58 of it and single points at 0.47; no background, no read noise, fixed random seed. Richardson–Lucy[24] runs live in your browser from a flat start. “Matched” deconvolves with the same point spread function that blurred the data; “20 % too narrow / too wide” uses that profile scaled radially by 0.8 or 1.2 (FWHM 150 or 224 nm), the kind of error a theoretical PSF makes when the real one is degraded or the settings differ. Images are linear, each clipped at its own 99.8th percentile. Dip: 1 − minimum ÷ mean of the two peaks on the averaged profile; separation: distance between the two peaks; beading: coefficient of variation of the summed intensity across the pair, row by row along 2.8 µm (0.003 for the true object). The iteration curves are computed offline with an exact port of this code (same random numbers) and match it. A best case for the method (exact blur model, no background, sparse bright lines), not any vendor’s implementation.
The math, with its sources
About the model
The array figures show pixel reassignment alone: the 31 element images, each shifted back by α·d (α = 0.47) and summed, with no deconvolution step, so the 141 nm is the gain of reassignment without any processing on top.
The array model has no noise. The deconvolution model has shot noise only, no background, a two-dimensional object of thin bright lines and, unless you choose a mismatched one, exact knowledge of the point spread function: the best case for any deconvolution. Vendor figures (Airyscan 1.7×; LIGHTNING “down to 120 nm”) are the manufacturers’ statements.
At the microscope
Both modes need Nyquist sampling for the enhanced resolution: smaller pixels than a standard scan (module 05).
Keep the raw data: every reconstruction is an estimate, and the processing strength trades resolution against noise amplification, as the iteration slider shows.
Deconvolve with a point spread function that matches the objective, wavelength, pinhole and sample (a measured bead image or a validated theoretical one): a mismatch either stalls the gain or invents fine structure.
Airyscan’s gain in signal-to-noise is useful for dim samples even when resolution is not the goal.[17]
Where does pixel reassignment’s gain come from?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [10]Scientific Volume Imaging. “Nyquist rate” and “Nyquist calculator”, svi.nl.
- [16]Sheppard CJR. Super-resolution in confocal imaging. Optik 80:53–54, 1988; Müller CB, Enderlein J. Image scanning microscopy. Physical Review Letters 104:198101, 2010.
- [17]Huff J. The Airyscan detector from ZEISS: confocal imaging with improved signal-to-noise ratio and super-resolution. Nature Methods 12:i–ii, 2015 (manufacturer application note).
- [19]Schermelleh L, Ferrand A, Huser T, Eggeling C, Sauer M, Biehlmaier O, Drummen GPC. Super-resolution microscopy demystified. Nature Cell Biology 21:72–84, 2019. doi:10.1038/s41556-018-0251-8
- [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.
- [24]Richardson WH. Bayesian-based iterative method of image restoration. Journal of the Optical Society of America 62:55–59, 1972; Lucy LB. An iterative technique for the rectification of observed distributions. Astronomical Journal 79:745–754, 1974.