Resolution and the Airy disk
Rayleigh, Abbe and FWHM, and the dip between two point images as they move together, for widefield and confocal detection.
Two questions, two numbers
A lens cannot focus light to a point. Diffraction at the circular aperture of the objective spreads the image of a point into a central disk with faint rings, the Airy pattern; its first dark ring lies at 0.61 λ/NA. Lord Rayleigh asked when two points that each give off their own light can still be told apart. He proposed that they are “just resolved” when the center of one disk falls on the first dark ring of the other; the summed brightness midway between the peaks then dips to about 73.5 % of the peak brightness. Summing the two images is right only when the points do not interfere, and in fluorescence they don’t: each molecule emits independently of its neighbors. For two fluorescent points in a conventional image, Rayleigh’s 0.61 λ/NA, with λ the emission wavelength, is the classical criterion to use; it is a convention, not a hard wall.[3][8]
Ernst Abbe asked a different question: what is the finest repeating pattern, such as the lines of a grating, that the lens can carry into the image? In a conventional image of a specimen that gives off its own light, as a fluorescent one does, the answer is a period of λ/(2NA): finer patterns reach the image with no contrast at all, and patterns just coarser with very little. Abbe himself worked with illuminated specimens, where the light leaving neighboring points can interfere. With light arriving straight along the axis, the finest period is then λ/NA; with a condenser of aperture NAcond it is λ/(NA + NAcond), reaching λ/(2NA) only when the condenser’s aperture matches the objective’s, so that some of the light arrives as obliquely as the objective can accept. When the light leaving two points is coherent (nearly so with the condenser stopped far down), Rayleigh’s criterion does not carry over: whether they show a dip between them depends on the relative phase of their light. At Rayleigh’s separation, two points in phase show no dip at all, and two in opposite phase show the brightness between them falling to zero.[3][7][25]
The full width at half maximum (FWHM) of the point image is a third, practical descriptor. All of these numbers scale with λ/NA: resolution improves with shorter wavelength and higher NA. For a conventional image from an aberration-free lens, nothing else enters these numbers (magnification does not); noise and coarse pixels can only make the achieved resolution worse, and the pinhole (module 03) and image scanning (module 10) change the picture.[4]
In a confocal microscope, the excitation spot and the pinhole both shape the point image. Closing the pinhole toward a fraction of an Airy unit narrows it further, at a steep cost in signal (module 03): in our model from 187 nm at 1 AU to 145 nm at 0.1 AU, 1.3× narrower, and 1.4× narrower than the widefield image (209 nm). At 1 AU the lateral resolution is essentially that of the excitation spot.[4][5]
Same separation, three microscopes
Move two points together and compare three ways of detecting them: a widefield camera, a confocal pinhole at 1 Airy unit, and one closed to 0.2 AU. Each mode has its own resolution numbers; the sharpest one pays in light.
Only the 0.2 AU pinhole gives a clear dip (46%), with 13% of the 1 AU peak signal (8% of the total light from a scanned point, module 10): sharper, but the image would need about 7.5 times the exposure for the same signal-to-noise ratio.
COMPUTED · LIVE. Two equal, incoherent point sources in focus. Widefield: the emission point image (520 nm). Confocal: excitation (488 nm) × emission (520 nm) seen through a pinhole of 1 or 0.2 Airy units. Vectorial model[20], ideal objectives, index-matched samples (oil n 1.518, water n 1.333, air), pinholes in each objective’s own Airy unit (453, 529 and 846 nm at the sample), no noise; radial profiles in 10 nm steps, linearly interpolated, as in the beginner’s lesson. Dip: how far the brightness midway between the points falls below the peaks; 26.5 % is the dip of Rayleigh’s criterion for the textbook Airy pattern, applied here to each real point image (separations rounded up to the next nm). “A dip appears” is Sparrow’s criterion (Sparrow 1916, Astrophys J 44:76): the separation at which the midpoint stops being a maximum (zero curvature); the values shown are within 2 nm of that point on the exact profiles. Dips between 0.01 % and 0.3 % read “< 0.3%”. Images are scaled to their own peak, or, with the toggle, the two confocal images share one scale (same laser power and detector), while the widefield image keeps its own (a camera under different illumination); “light” is the in-focus peak signal of a point (the brightest pixel of its image) relative to a 1 AU pinhole at the same laser power (module 03), from the encircled energy of the point image inside each pinhole. Summed over the whole scanned image of the point, the 0.2 AU pinhole keeps 8 % of the 1 AU signal at NA 1.4 oil, against 13 % at the peak (module 10). The textbook values (0.61 λ/NA, λ/(2NA)) are scalar widefield formulas and do not change with the pinhole; the model’s own distances for each mode are in the table and on the chart.
The math, with its sources
About the model
Points are ideal emitters of equal brightness; the model ignores noise, aberrations, scattering and finite pixel size. In a real image, a shallow dip is lost unless the signal-to-noise ratio is high and the sampling is fine enough (module 05).
At the microscope
Use the highest-NA objective that suits the sample, with the immersion it was designed for, and #1.5 coverslips (0.17 mm) for high-NA objectives.
Measure your own point spread function with sub-resolution beads mounted like your sample: it is the honest number for your system. Use beads well below the resolution (a bead’s own size widens the measured FWHM), as in the protocol of Cole, Jinadasa & Brown 2011 (Nat Protoc 6:1929).
Which of these numbers is the finest grating period a lens can transmit?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [3]Born M, Wolf E. Principles of Optics, 6th ed. Pergamon, 1980. §8.5.2 (Airy pattern, encircled energy), §8.6.2 (Rayleigh criterion), §8.6.3 (image formation in the microscope: incoherent illumination; coherent illumination, Abbe’s theory) and §8.8 (three-dimensional light distribution near focus).
- [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
- [5]Wilson T. Resolution and optical sectioning in the confocal microscope. Journal of Microscopy 244:113–121, 2011. doi:10.1111/j.1365-2818.2011.03549.x
- [7]Abbe E. Beiträge zur Theorie des Mikroskops und der mikroskopischen Wahrnehmung. Archiv für mikroskopische Anatomie 9:413–468, 1873.
- [8]Rayleigh, Lord. Investigations in optics, with special reference to the spectroscope. Philosophical Magazine (5) 8:261–274, 1879.
- [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.
- [25]Goodman JW. Introduction to Fourier Optics, 4th ed. W. H. Freeman, 2017. §7.5 (comparison of coherent and incoherent imaging; §7.5.2, two-point resolution).