The pinhole: trading signal for sectioning
In-focus signal, out-of-focus rejection, and the axial and lateral response, computed with a vectorial model for any pinhole size and three objectives.
What the pinhole size buys and costs
Pinhole size is quoted in Airy units: 1 AU = 1.22 λem/NA, the diameter of the Airy disk of the emission light in the sample plane. The physical pinhole is this value times the magnification between sample and pinhole, which the software handles.[3][4]
The fraction of an in-focus point’s light that passes the pinhole is the encircled energy of its image within the pinhole radius. The scalar textbook result at 1 AU is 84 %.[3] Our full vectorial calculation for a 1.4 NA oil objective (vectorial model[20]; randomly oriented emitters, index-matched, aberration-free) gives 76 %: at high aperture the image of an emitter is wider than the Airy pattern, partly because the steep rays carry more weight, and partly because emitters whose dipole points along the axis image as a ring rather than a spot, so more of the energy falls outside 1 AU.[26] The difference matters when you compare pinhole settings quantitatively.
Closing the pinhole below 1 AU narrows the response laterally and axially, but the signal falls fast; opening it beyond 1 AU adds little signal and a lot of out-of-focus light. The optical section of a thin fluorescent sheet grows almost linearly with pinhole size above 1 AU, which the Zeiss slice-thickness formula captures with its geometric term.[4][5]
Change the pinhole, watch the trade
Slide the pinhole from 0.1 to 4 Airy units for three objectives. The detector-plane pictures show what the pinhole lets through from a point in focus and from a point out of focus; the curves show the whole trade at once, and the readouts give the numbers for your setting.
The usual compromise: 76% of the in-focus light passes and the section is 501 nm for this objective. A point 1.0 µm out of focus gets 4.6% of its light through.
Vectorial model[20], 488 nm excitation, 520 nm emission, aberration-free objectives, index-matched sample: NA 0.75 air (1 AU = 846 nm), NA 1.2 water (529 nm), NA 1.4 oil (453 nm). In-focus light that passes: encircled energy of the emission point image inside the pinhole, as a share of all the light the objective collects from that point. Out-of-focus point: the same for a point z µm from focus, assuming it emits as much as the point in focus (the detection side only; the weaker excitation out of focus is included in the axial FWHM and the section, not here). Section: FWHM of the signal from a thin fluorescent sheet scanned through focus. Axial and lateral FWHM: of the confocal response to a point object. Every 0.05 AU step is computed exactly for the pass fractions, the curves and the axial FWHM; the lateral FWHM and the section are interpolated between 18 computed pinhole sizes (checked against exact values at 12 intermediate sizes: within 1.5 nm for oil, 2.2 nm for water, 5.5 nm for air). Detector-plane pictures: the computed emission point image h(r, z) on a 6 × 6 µm field at the sample scale, both on one logarithmic brightness scale spanning three decades below the in-focus peak so faint light stays visible (inferno colour map); the circle is the pinhole, au × 1 AU in diameter, and the plate outside it is shaded.
The math, with its sources
About the model
A single point or sheet, no aberrations, scattering or noise. Real point spread functions, measured with sub-resolution beads, are always somewhat wider. The scalar (84 %) and vectorial (76 %) in-focus values differ because the scalar model ignores polarization and the high-angle apodization.
All curves are for an index-matched sample at the focus. Deeper in an aqueous sample under an oil objective, spherical aberration thickens the section and lowers the signal (module 06).
At the microscope
The acquisition software expresses the pinhole in Airy units, computed for the current objective and emission band. Record the value with every image, since it sets both the section thickness and the signal.
For dim samples, opening to 1.2–1.5 AU collects more light, mostly from planes just above and below focus: in the model the in-focus signal rises only 3–9 % while the section grows 10–30 % (see the curves above), so it is a trade worth making only when sectioning matters less than signal. Below 0.5 AU the signal falls faster than the resolution improves, unless the detector is an array (module 10).
At NA 1.4, you close the pinhole from 1 AU to 0.5 AU. What do you get?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [3]Born M, Wolf E. Principles of Optics, 6th ed. Pergamon, 1980. §8.5.2 (Airy pattern, encircled energy), §8.6.2 (Rayleigh criterion), §8.6.3 (image formation in the microscope: incoherent illumination; coherent illumination, Abbe’s theory) and §8.8 (three-dimensional light distribution near focus).
- [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
- [5]Wilson T. Resolution and optical sectioning in the confocal microscope. Journal of Microscopy 244:113–121, 2011. doi:10.1111/j.1365-2818.2011.03549.x
- [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.