The confocal principle
One point lit, one point seen: the conjugate pinhole, scanning and descanning, and why the confocal point spread function is a product.
The mirror has moved the spot 25 µm across the sample. The fluorescence comes back over the same mirror, which undoes the tilt, so its image stays in the centre of the pinhole: the detector sees the spot wherever it is.
One point lit, one point seen
A widefield fluorescence microscope lights the whole field and collects light from every depth at once. Marvin Minsky’s 1957 design removed the out-of-focus haze with two ideas: light a single point, and put a small aperture, the pinhole, in a plane conjugate to the focal plane in front of the detector.[2]
Because only one point is observed at a time, the image is built by scanning. Minsky moved the specimen; today’s laser scanning confocals sweep the focused spot across the sample with two galvanometer mirrors, and the fluorescence returns through the same mirrors (descanned detection), so its image stays fixed on the pinhole. A photomultiplier or hybrid detector records one value per pixel.[4]
The confocal point spread function is the product of where light is delivered and where it is collected: the excitation PSF times the detection PSF seen through the pinhole. The product is always narrower than either factor, but it gains most when both are narrow: with a very small pinhole it is about 1.4 times narrower than the widefield image. With a pinhole of 1 AU or more the detection factor is broad, lateral resolution is set mainly by the excitation spot, and the pinhole’s job is depth discrimination.[4][5]
Every plane sends the same total light
Energy conservation says that the light from a point emitter crosses every plane between the sample and the detector with the same total power. A camera integrates over the whole image plane, so an out-of-focus plane in the sample still sends all its light to the camera: blurred, but not dimmer. That is why widefield images of thick samples carry haze, and why no setting can remove it optically.[3]
In a confocal, light from an out-of-focus plane arrives at the pinhole spread out and is mostly blocked. A thin sheet absorbs the same total laser light at every depth, but out of focus the glow it gives off is spread over a wide area, so little of it lines up with the pinhole. The total signal from a thin fluorescent sheet therefore falls off with its distance from focus. The width of that falloff is the optical section, plotted in the model below for each of 18 pinhole sizes.[4][5]
Illumination × detection = confocal
Slide the pinhole and watch the product: the laser focus stays fixed, the detection hourglass narrows or widens with the pinhole, and their product, the confocal point spread function, follows; click a map to read the multiplication at any point. The chart shows what makes the difference: how much light a thin fluorescent sheet still sends when it is out of focus.
At 1 AU the product is compact in x, y and z: a thin sheet 2 µm from focus sends only 0.82% of its in-focus signal. Laterally it is set mainly by the laser focus (187 nm against 197 nm for the laser alone); closing the pinhole to 0.2 AU narrows it by 21 % (to 147 nm), but then only 10.0% of the in-focus light passes instead of 75.9%.
COMPUTED · LIVE. Vectorial (Richards–Wolf) model[20] of an aberration-free 1.4 NA oil objective, index-matched sample, 488 nm excitation, 520 nm emission, circularly polarized light and freely rotating emitters. Left: the laser focus. Middle: the emission point image seen through a pinhole of the chosen size (a 2D convolution in every plane, the disc supersampled 4 × 4). Right: their product, the confocal point spread function, computed pixel by pixel in your browser from the two maps; the detection point spread function of isotropic emitters has the same vectorial form as the illumination one, by reciprocity. x–z sections 2.4 × 4 µm on a 20 nm grid, square-root brightness, each map scaled to its own peak; FWHMs and all percentages are computed on linear values. Sheet signal: total fluorescence from a uniform thin sheet at distance z from focus, relative to the same sheet in focus; widefield stays at 100 % because every plane passes the same total power (our model: within 0.23 %); far from focus the confocal sheet signal falls off as 1/z² (Sheppard & Wilson 1978, Opt Lett 3:115; Wilson 1989, J Microsc 154:143). The 18 pinhole sizes are each computed exactly; nothing is interpolated between sizes. The last stop removes the pinhole: the detection factor is then the same everywhere and the product is the laser focus alone. Probe: the two maps read at one point on linear values, each 1 at the focus (bilinear between the 20 nm grid points), so the detection value is relative to an in-focus point: 6.1 % at 1 µm and 1 AU is the 4.6 % of a point’s light that passes the pinhole in §2, divided by the 76 % that passes in focus.
The math, with its sources
About the model
Aberration-free optics, index-matched sample, no scattering. The laser is assumed to fill the objective’s back aperture uniformly and to be circularly polarized. Real lasers are usually linearly polarized, which at NA 1.4 stretches the laser focus along the polarization (in this model about 234 × 168 nm instead of 197 nm round), and a Gaussian beam that underfills the aperture widens it a little. The x–z maps are shown on a square-root brightness scale so the faint cones and side lobes are visible; the numbers are computed on the linear values.
At the microscope
Galvanometer mirrors sweep the spot continuously along each line, and the detector signal is summed into pixels (Pologruto, Sabatini & Svoboda 2003, BioMed Eng OnLine 2:13); resonant scanners sweep much faster and trade dwell time for speed. Either way the pinhole sees a stationary image, because detection is descanned.
Sectioning matters most in thick samples: for very thin specimens, such as the flat edges of cultured cells, a widefield image can be nearly as sharp and much faster (Basics, “Widefield or confocal?”).
Why does a widefield image of a thick sample carry haze that no setting can remove?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [2]Minsky M. Microscopy apparatus. US Patent 3,013,467, filed 7 November 1957, issued 19 December 1961.
- [3]Born M, Wolf E. Principles of Optics, 6th ed. Pergamon, 1980. §8.5.2 (Airy pattern, encircled energy), §8.6.2 (Rayleigh criterion), §8.6.3 (image formation in the microscope: incoherent illumination; coherent illumination, Abbe’s theory) and §8.8 (three-dimensional light distribution near focus).
- [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
- [5]Wilson T. Resolution and optical sectioning in the confocal microscope. Journal of Microscopy 244:113–121, 2011. doi:10.1111/j.1365-2818.2011.03549.x
- [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.