Module 01 · Chapter I · Light and the instrument

Fluorescence: where the photons come from

Absorption, the Stokes shift, quantum yield and lifetime, and the triplet route to bleaching, with measured spectra for six fluorescent proteins and any laser line.

Rendered close-up of green fluorescence returning from the lit point in the sample into the objective RENDER
Fluorescence from the lit spot travels back into the objective. From the Core’s animation of the light path.
1 · The physics

From absorbed photon to emitted photon

A fluorophore absorbs a photon and is promoted from its electronic ground state, S₀, to an excited singlet state. Within picoseconds it relaxes to the lowest vibrational level of S₁, losing some energy as heat. A few nanoseconds later it returns to S₀, and the remaining energy leaves as a fluorescence photon. Because energy is lost on both sides of the cycle, the emitted photon carries less energy, and so has a longer wavelength, than the absorbed one. That shift between absorption and emission is the Stokes shift, and it is what lets a dichroic mirror and an emission filter separate the faint fluorescence from the bright excitation light.[1]

Two numbers describe how much light a fluorophore returns. The extinction coefficient ε says how strongly it absorbs at its peak; the quantum yield Φ is the fraction of absorbed photons that come back out as fluorescence. Their product, ε·Φ, is the usual brightness figure. The lifetime τ, the average time spent in S₁, is typically 1 to 10 ns. Because lifetimes differ between molecules, lifetime imaging can separate two dyes of the same color; because τ also depends on the surroundings, it can report on them.[1]

From S₁ a molecule can also cross into a long-lived triplet state, T₁. Triplet molecules react readily with oxygen and with their neighbors, and those reactions are a main route to photobleaching (module 08).[1][9]

S₀S₁T₁ absorption~10⁻¹⁵ s relaxation, ~10⁻¹² s fluorescenceafter ~1–10 ns intersystem crossing reactions with O₂ andneighbors: bleaching
Simplified Jablonski diagram after Lakowicz. Levels and arrow lengths are schematic; time scales are typical orders of magnitude.[1]
2 · Live model

Six fluorescent proteins, any laser line, your detection band

Pick a fluorophore, tune the excitation laser, and set the detection band. The chart shows what the laser excites and what the band collects; the readouts turn that into an expected signal, with every constant from FPbase, and the table repeats the calculation for all six proteins in the same channel.

Measured spectra, each scaled to its own peak (FPbase, 1 nm steps). excitation · filled: emission · brighter fill: emission inside the detection band Stokes shift 19 nm 488 nm · 100% of peak detect 500–550 nm 400450500550600650700750800050100wavelength (nm)% of peak
Photon energies, E = hc/λ, to scale 0123energy (eV) S₀S₁ 2.54 eV488 nm in 2.45 eV507 nm out 0.10 eV 3.7% as heat
All six in this channel: laser 488 nm, band 500–550 nm
ProteinExcitedCollectedSignal vs reference
mTurquoise20.7%38%0.003×
EGFP100%72%1.00×
mNeonGreen62%80%1.90×
mScarlet15%0.1%< 0.001×
mCherry7.6%0.0%< 0.001×
miRFP670—0.0%no data
Reference 1.00× = EGFP, 488 nm, 500–550 nm; bars scaled to the largest value. No measured excitation data at 488 nm for miRFP670 (520–713 nm).
Fluorophore
Excitation laser488 nm (tunable)
405 and 448 nm diodes · white-light laser tunable 485–790 nm (the slider steps only through these lines)
Laser presets
Detection band starts at500 nm
Detection band ends at550 nm
50 nm wide, just above the laser line, at the emission peak
Excitation / emission peak (FPbase)488 / 507 nm
Stokes shift19 nm · 0.10 eV
ε (M⁻¹cm⁻¹) × Φ55,900 × 0.60
Brightness ε·Φ ÷ 1000 · lifetime τ33.5 · 2.6 ns
Excitation at the laser, of its peak100%
Emission collected by the band72%
Expected signal vs EGFP at 488 nm, 500–550 nm1.00× ε·Φ 33.5 × excitation 100% × collected 72%, ÷ 33.5 × 100% × 72% for the reference (bar: 4× = full)

A good match: EGFP absorbs 100% as well as at its peak and the band collects 72% of its emission, for 1.00× the reference signal.

Spectra: FPbase[21] 1 nm data (fpbase.org/spectra_csv, CC BY-SA 4.0), each curve scaled to its own peak; the mTurquoise2 excitation curve is its absorption spectrum, which is proportional to the excitation spectrum when the quantum yield does not depend on wavelength. Published values used, with the wavelength range of each excitation / emission curve as used here (FPbase data below 380 nm omitted): mTurquoise2: 434/474 nm, ε 30,000 M⁻¹cm⁻¹, Φ 0.93, τ 4.0 ns, pKa 3.1 (FPbase 7AV5G, spectra 1745/1746, measured 380–495 / 441–650 nm); EGFP: 488/507 nm, ε 55,900 M⁻¹cm⁻¹, Φ 0.60, τ 2.6 ns, pKa 6.0 (FPbase R9NL8, spectra 17/18, measured 380–700 / 463–700 nm); mNeonGreen: 506/517 nm, ε 116,000 M⁻¹cm⁻¹, Φ 0.80, τ 3.1 ns, pKa 5.7 (FPbase ZRKRV, spectra 101/102, measured 380–600 / 472–748 nm); mScarlet: 569/594 nm, ε 100,000 M⁻¹cm⁻¹, Φ 0.70, τ 3.9 ns, pKa 5.3 (FPbase FVS3D, spectra 119/120, measured 384–644 / 538–773 nm); mCherry: 587/610 nm, ε 72,000 M⁻¹cm⁻¹, Φ 0.22, τ 1.4 ns, pKa 4.5 (FPbase ZERB6, spectra 79/80, measured 380–650 / 550–800 nm); miRFP670: 642/670 nm, ε 87,400 M⁻¹cm⁻¹, Φ 0.14, τ 1.42 ns, pKa 4.5 (FPbase DP999, spectra 1604/1606, measured 520–713 / 606–812 nm). Excitation at the laser = ex(λlaser)/ex(peak), read from the curve; where the laser falls outside the measured range the model reports "no data" rather than zero. Collected fraction = emission summed over the band ÷ emission summed over the whole measured curve, 1 nm steps. Expected signal ∝ ε·Φ × excitation × collected fraction, for equal photon flux (photons per second; at equal power a longer wavelength delivers proportionally more photons), equal amounts of mature protein, and a detector of equal efficiency at every wavelength; reference: EGFP, 488 nm, 500–550 nm (ε·Φ = 33.5 × 10³, excitation 100 %, collected 72 %). Peaks and Stokes shifts are the FPbase values; the measured curves peak within 1–4 nm of them. Photon energy E = hc/λ with hc = 1239.84 eV·nm; the Jablonski levels are drawn to that scale, with S₁ placed at the energy of the emission-peak photon (so the whole Stokes loss is drawn on the S₁ side for simplicity; in reality part of it happens in S₀ after emission) and vibrational spacings schematic. A laser photon with less energy than that is marked “anti-Stokes” (only molecules with extra thermal energy absorb it); beyond the measured excitation curve and more than 0.25 eV short (about 10 kT at room temperature, so fewer than 1 in 10,000 molecules have the extra energy), it is marked as unable to excite the molecule on its own (two photons can, module 09). Laser lines as on the Core’s Leica Stellaris: 405 and 448 nm diodes plus a white-light laser tunable from 485 to 790 nm; the colour strip under the axis is an approximate rendering of wavelength.

Formulas used in this module

The math, with its sources

Photon energy[1]
E = hc / λ
With hc = 1239.84 eV·nm: a 488 nm photon carries 2.54 eV; one at 507 nm, 2.45 eV. The difference is lost as heat.
Stokes shift[1][21]
Δλ = λem,max − λex,max
In nm. EGFP: 507 − 488 = 19 nm; mCherry: 610 − 587 = 23 nm; mTurquoise2: 474 − 434 = 40 nm.
Quantum yield and lifetime[1]
Φ = kr / (kr + knr),τ = 1 / (kr + knr)
kr: radiative rate; knr: sum of non-radiative rates (internal conversion, intersystem crossing, quenching). Anything that adds a non-radiative path lowers both Φ and τ.
Brightness[1][21]
B ∝ ε(λlaser) · Φ,ε(λlaser) = εmax · ex(λlaser) / ex(λmax)
ε at the excitation wavelength actually used, scaled from the peak value with the normalized excitation spectrum. Detected signal also scales with laser power, collection efficiency, filters and detector quantum efficiency.
Signal in a detection channel[1][21]
S ∝ ε(λlaser) · Φ · C,C = ∫band em(λ) dλ / ∫ em(λ) dλ
C: the fraction of the emission spectrum that falls inside the detection band. For EGFP a 500–550 nm band collects 72 % of the emission, a 600–650 nm band 2 %; the live model evaluates this for any band.

About the model

Spectra are measured curves from FPbase at 1 nm resolution; the model reads them directly, with no fitting. Brightness uses the published ε and Φ; real signal also depends on how much protein is expressed and matured, on pH (the pKa values are listed in the fine print: EGFP 6.0, mNeonGreen 5.7, mTurquoise2 3.1) and on the optics and detector. Comparing proteins assumes equal amounts of mature protein, which fusions rarely give.

At the microscope

Ask which laser lines your fluorophores need before booking: a dye excited off its band needs far more power for the same signal, and the extra light adds phototoxicity, autofluorescence and excitation of other dyes (module 08). A tunable laser can sit on the excitation peak, which the model shows as extra excitation for the same photon flux.

Lifetime differences (EGFP 2.6 ns, mCherry 1.4 ns, mScarlet 3.9 ns) can separate fluorophores that look alike in color, on systems equipped for lifetime imaging.

Check yourself

EGFP absorbs best at 488 nm but emits around 507 nm. Why is the emitted light redder?

References

Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.

  1. [1]Lakowicz JR. Principles of Fluorescence Spectroscopy, 3rd ed. Springer, 2006. Chapter 1 (Jablonski diagram, Stokes shift, quantum yield, lifetime).
  2. [9]Pawley JB (ed.). Handbook of Biological Confocal Microscopy, 3rd ed. Springer, 2006. Chapter 4 (Pawley, “Points, pixels, and gray levels”) and Chapter 39 (Diaspro et al., “Photobleaching”).
  3. [21]Lambert TJ. FPbase: a community-editable fluorescent protein database. Nature Methods 16:277–278, 2019. Spectra from fpbase.org (CC BY-SA 4.0). doi:10.1038/s41592-019-0352-8