Why cells glow
Fluorescent dyes take in light of one color and give back light of a longer wavelength: blue in, green out. The microscope keeps the two apart, so only the glow reaches the detector.
Light in, a different color out
A fluorescent molecule (a fluorophore, or dye) absorbs light of one color. A few billionths of a second later, on average, it gives out light of its own. A little energy is lost as heat along the way, so the light that comes out is shifted to longer wavelengths than the light that went in: blue in, green out; green in, orange out.
Microscopists attach these dyes to what they want to see: an antibody that sticks to one protein, a stain that binds DNA, or a fluorescent protein such as GFP that the cell makes itself.
Every dye has two curves: which colors it absorbs (excitation) and which colors it gives off (emission). The gap between them (the Stokes shift) is what lets the microscope separate the faint glow from the much brighter light used to create it.
Blue in, green out: 4% longer waves, 4% less energy at the peak
Tune the laser, set the band
These are the measured curves for EGFP, a green fluorescent protein. Tune the laser that lights it and set the band of colors that reaches the detector, then see what the detector records.
A good match: the laser excites EGFP at 100% of its peak and your band collects 72% of its glow while blocking the laser, for 1.00× the reference signal.
COMPUTED / SIMULATED · Measured EGFP spectra from FPbase (fpbase.org), 1 nm steps, each scaled to its own peak, the same data and definitions as module 01. “Excitation” is the excitation curve at the laser wavelength, as a share of its peak (a relative cross-section, not the share of laser light absorbed). “Light inside your band” is the share of the emission curve between the band’s edges (380–800 nm total). Signal = excitation × light inside the band, relative to 488 nm and 500–550 nm, for equal photon flux (at equal power a 405 nm beam carries 0.83× the photons of a 488 nm beam). The laser slider snaps to the Core’s Leica STELLARIS lines (405 and 448 nm diodes, white-light laser 485–790 nm). The cell is the site’s simulated cell (microtubules, confocal at 1 AU, NA 1.4 oil, recorded with 80 nm pixels) on one brightness scale, with photon noise from a fixed seed: about 10 photons per pixel at the brightest microtubules at the reference setting (the Basics pyramid’s middle setting: 80 nm pixels, 4 µs per pixel). When the band includes the laser line, or one of its edges comes closer than 5 nm to it (a page convention: real filters need a few nanometers to block the laser), the detector is drawn saturated; the actual scatter level is not modeled. Detector efficiency is taken as flat across the band. Screen colors are approximate: pure spectral colors lie outside what screens can show, so each wavelength is drawn as the most saturated screen color of the same hue, at full brightness (CIE 1931 → sRGB). The objective catches about 31% of the glow: the share of light from molecules glowing equally in all directions that falls inside the cone of an NA 1.4 oil objective, (1 − cos α)/2 with α = 67.3°, for a sample mounted in a medium matched to the oil (as in the simulated cell); it ignores the emission pattern of single fixed molecules, losses inside the objective, and, for cells in water, the change of the cone at the coverslip.
Remember
- 1A fluorescent dye absorbs one color and gives off a longer-wavelength color.
- 2A filter in front of the detector blocks the laser and passes the dye’s glow.
- 3Match the laser to the dye’s excitation, and the detector band to its emission.