Microscopy basics · Lesson 2 of 8 · about 5 minutes

One spot at a time

A point-scanning confocal sweeps one tiny spot across the sample, line by line, and records how much light comes back at each point. The image is built one pixel at a time.

Trades: Detail · Speed Confocal only Go deeper: the advanced module on the confocal principle →
Rendered view of the two small scan mirrors that steer the laser beam across the sample RENDER
RENDER · Two small mirrors tilt to steer the laser spot across the sample, line by line. From the Core’s animation of the light path.
1 · The idea

A camera takes it all at once. A confocal reads it point by point.

A widefield microscope lights the whole field of view and records all pixels in one exposure with a camera.

A point-scanning confocal works differently. It focuses the laser to a spot about 0.2 µm wide inside the sample and measures how much light comes back from that one spot with a very sensitive detector. Then two small mirrors move the spot: one sweeps it along a line, the other steps it to the next line, like reading a page. The computer puts the measurements together into the image.

Lighting one spot at a time is what makes the pinhole trick in lesson 3 possible. The cost is time: every pixel is a separate measurement.

One image = 262,144 separate measurements

SIMULATED · 41 × 41 µm
COMPUTED · TO SCALE197 nmat half heightexcitation spot (488 nm, NA 1.4 oil)one pixel = 80 nmfast axis: the spot sweeps on, 2 µs per pixelflyback to the start of the next line,80 nm lower (the laser is often switched off)
MODEL · TIMINGone pixel · 2 µsone line = 512 pixels · 1.02 ms on the pixels, up to ≈ 1.3 ms with the mirror’s turnaroundone image = 512 lines · 0.52 s on the pixels, up to ≈ 0.66 s in all1 µs10 µs100 µs1 ms10 ms100 ms1 sturnaround and flyback, at most 25% of the pixel timelog scale: each step = 10×
SIMULATED · COMPUTED · MODEL · The simulated cell (microtubules and DNA, confocal at 1 AU, NA 1.4 oil, 488/520 nm) recorded by a point scanner with 512 pixels per line over a 41 µm field (80 nm pixels) at 2 µs per pixel: about 5 photons per pixel at the brightest microtubules (the Basics pyramid’s photon scale), with photon noise. The spot is the computed excitation focus (vectorial model, FWHM 197 nm), drawn to scale over the pixels. Timing: pixels × time per pixel, plus up to 25% for the mirror’s turnaround and flyback at each line end (an upper estimate from Zeiss’s LSM 800 figure; details under the try-it); the frame flyback is ignored. Pixels are time bins of a continuous sweep, not stops.
2 · Try it

Watch an image being built

Choose how long the spot takes to cross each pixel, how many pixels make up each line, and whether each line is averaged. The times come from a timing model for a galvo scanner; the animation draws every line at the speed stated above the image.

Playing: every one of the 512 lines, shown 6.1× slower than real time.
SIMULATED · LIVE · 41 × 41 µm
Sample drift · sideways, while it is scannednone
The sample holds still.
Each line is recorded later than the one above it: the strip on the left shows when (green at the start of the image, orange at the end). Below the current line you still see the previous image, dimmed. The square at the bottom right magnifies a 3.84 µm detail to show your pixels.
Start from · typical settings
Quick enough to follow slowly moving cells; still grainy.
Time per pixel · how long the spot takes to cross each pixel2 µs
Pixels per line
Line averaging
Pixel size80 nm262,144 measurements
Photons per pixel≈ 5brightest microtubules · grainy
Time per image≤ 655 ms524 ms on the pixels
Images per second≥ 1.51.9 on the pixels alone
MODEL · LIVETime per image, on a log scale · marks: fastest settings at 512 × 512
10 ms100 ms1 s10 s100 s Zeiss LSM 800, fastest galvo: 8 images/s Nikon A1R, resonant scanner: 30 images/s green: pixels alone · white mark: with turnaround your image · 655 ms

About 1.5 images per second with 5 photons per pixel at the brightest microtubules: quick enough for slowly moving live cells, but still grainy. Doubling the time per pixel or scanning each line twice (2× line averaging) doubles the light and the time.

SIMULATED · MODEL · The simulated cell (microtubules and DNA, confocal at 1 AU, NA 1.4 oil, 488/520 nm), recorded the way a point scanner does: each pixel adds up the light while the spot sweeps across its width, line after line, over a 41 × 41 µm field (the smooth DNA channel is stored at 320 nm and interpolated). Photons per pixel ∝ time per pixel × averages: about 10 at 4 µs for the brightest microtubules (the Basics pyramid’s scale, laser power fixed), with photon noise from a fixed seed so only your settings change the grain; each image is shown on its own expected brightness, so the grain changes, not the brightness. Time per image = pixels × time per pixel × averages, plus the mirror’s turnaround and flyback at each line end, added as 25% of the pixel time, the most Zeiss’s figure for the LSM 800 allows (more than 80% of the scan time is used for data), so times “with turnaround” are upper estimates (a galvo’s turnaround takes about the same time on every line, so for long lines, slow or many pixels, its real share is much smaller than 25%). Lines are limited to about 4,096 per second (from the LSM 800’s 8 images/s at 512 × 512 and 64 images/s at 512 × 64; the brochure does not say whether that fastest setting scans in both directions); at that limit the time per pixel is set by the sweep, 80% of 1/4,096 s ÷ pixels per line (0.76 µs at 256 pixels). The slider stops at 0.5 µs, so this model’s fastest 512 × 512 image takes 164 ms; 8 images/s needs a shorter time per pixel (about 0.38 µs in this model). The frame flyback is ignored. Resonant scanners sweep thousands of lines per second (Nikon A1R: 7.8 kHz, 30 images/s at 512 × 512; resonant scanners are also an option on the Leica STELLARIS, but the Core’s STELLARIS has the galvanometric scanner), so their time per pixel is much shorter. Because they swing sinusoidally, it also varies along the line: at 7.8 kHz, with 512 pixels over the central 90% of the swing (our model), from about 0.07 µs at the center to 0.16 µs at the edges. Drift: each line samples the cell shifted sideways by the drift speed × the time at which the line is recorded (linear interpolation of the 80 nm scan; the simulated data reach 6 µm beyond each side of the field, so the drift is capped at 6 µm per image).

Remember

  1. 1A point-scanning confocal measures one spot at a time and builds the image line by line.
  2. 2Time per image ≈ number of pixels × time per pixel × averages, plus the mirrors’ turnaround at the end of each line.
  3. 3Lighting one spot at a time is what lets the pinhole reject out-of-focus light (lesson 3).
Check yourself

A confocal records 1024 × 1024 pixels at 2 µs per pixel. About how long does one image take?

Training

These pages are purely to improve your knowledge: practical training on the instrument is not only recommended but mandatory.