Z-stacks and axial sampling
Optical section, z step and slice count for your objective, immersion, laser line and pinhole, computed from the vectorial model, with a simulated stack and what index mismatch does to it.
Axial resolution, sectioning and the step size
The axial response of a confocal is always worse than the lateral one. For a point object with the pinhole at 1 AU or more, the Zeiss expression gives an axial FWHM of 0.88 λexc/(n − √(n² − NA²)), roughly 2.5 times the lateral FWHM for a high-NA oil objective. At low NA it reduces to about 1.77 n λ/NA²: axial resolution depends on the square of the aperture.[4]
For an extended fluorescent sheet the measured section is thicker than for a point, because light from the whole plane reaches the pinhole: about 500 nm for a 1 AU pinhole at NA 1.4 in our model, against 391 nm for a point. Nyquist applies along z as it does laterally: step by no more than half the section, or use the bandwidth rule Δz = λexc/(4n(1 − cos α)) ≈ 130 nm if the stack will be deconvolved.[4][10]
A caution: when the immersion and the sample have different refractive indices, the focus moves through the sample by a different distance than the objective moves, and spherical aberration broadens and dims the axial response with depth. With an oil objective focusing into an aqueous sample, the stack is elongated along z and blurred unless corrected.[13][23]
Your z step, slice count and sample
Choose the objective, laser line and pinhole, then move the z step: the upper view shows the object as the optics blur it, with every focal plane the stack records; the lower view shows what the stack stores. The readouts are your numbers. Then mount the sample in water under an oil objective.
Step 1.2× the practical rule: beads between planes come out dimmer or in one plane only, and the tilted filament breaks into steps that interpolation smooths but cannot correct.
Vectorial confocal model[20] (Richards–Wolf, 488 nm excitation, 520 nm emission, circular polarization, pinhole integrated plane by plane), computed for nine objectives (air 0.40, 0.75, 0.95; water 1.00, 1.10, 1.20; oil 1.30, 1.40, 1.45) at 13 pinhole sizes and interpolated in between; the x–z views use maps at 0.2, 0.5, 1, 2 and 4 AU, interpolated. Other laser lines scale every distance, including depths, by λexc/488 with the emission at 1.066 λexc, which is exact for this model. Section and axial FWHM: full widths at half maximum of the thin-sheet and bead responses, for an index-matched sample; the step rules use these best-case values. Beads and the filament, which lies in the x–z plane, are convolved with the x–z point spread function; the 3 µm wide sheet uses the point spread function integrated across y. Recorded planes: the image sampled at each focal position with pixels of half the lateral FWHM, Poisson noise with 150 photons in the brightest pixel of the index-matched 1 AU scene for the chosen objective and line; other pinholes and the aqueous sample scale with the light that reaches the detector. Aqueous sample (n 1.333): the same model extended to focusing through a flat interface at the coverslip surface (Fresnel transmission; phase k0(nsample z cos θ2 − nimm d cos θ1) for a stage travel d; rays beyond the critical angle evanescent), for excitation and, by reciprocity, detection.[13] Focus = peak of a bead’s response; bead signal relative to a bead 0.5 µm above the coverslip. The x–z views use the aberrated point spread function computed for a focus 3 µm into the sample across the whole 6 µm window. Scattering, absorption and correction collars are not modeled. Time and dose assume the same frame time and laser power per slice.
The math, with its sources
About the model
The live model treats the sample as clear and the objective as perfect: an index-matched sample keeps the same point spread function at every depth. In an aqueous sample under an oil or air objective it computes the focal shift, the broadening of the section and the loss of signal with depth from the flat interface alone, the effects the Diel et al. tutorial explains how to measure and correct. Scattering adds to both losses in thick tissue.[23]
At the microscope
Match the immersion to the sample where you can: water objectives for live or aqueous samples, oil for samples mounted in a medium of index 1.52.
Set the top and bottom of the stack just around the region you need: the dose grows with the number of slices, and so does the time.
You image an aqueous sample with an oil objective and a nominal 0.3 µm z step. What happens to the stack?
References
Numbered as in the full list on the track page. Figures from a manufacturer rather than the literature are identified as such where they are used.
- [4]Wilhelm S, Gröbler B, Gluch M, Heinz H. Confocal Laser Scanning Microscopy: Principles. Carl Zeiss Microscopy, Jena, 2010. Part 1, equations (2)–(8); Part 2 (Nyquist theorem, pixel size); Part 3, ‘Details’, p. IV (fluorescence, photobleaching quantum efficiency).
- [10]Scientific Volume Imaging. “Nyquist rate” and “Nyquist calculator”, svi.nl.
- [13]Hell S, Reiner G, Cremer C, Stelzer EHK. Aberrations in confocal fluorescence microscopy induced by mismatches in refractive index. Journal of Microscopy 169:391–405, 1993.
- [20]Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proceedings of the Royal Society of London A 253:358–379, 1959.
- [23]Diel EE, Lichtman JW, Richardson DS. Tutorial: avoiding and correcting sample-induced spherical aberration artifacts in 3D fluorescence microscopy. Nature Protocols 15:2773–2784, 2020. doi:10.1038/s41596-020-0360-2