Spectra, bleed-through and unmixing
Detection bands, laser lines, sequential scanning and linear unmixing, computed from the measured spectra of EGFP and mCherry.
Crosstalk, and three ways to remove it
Fluorophore spectra are broad, and emission spectra have long tails toward the red. When two dyes are imaged together, some light from the shorter-wavelength dye falls into the detection band of the longer-wavelength one. That bleed-through (crosstalk) can masquerade as colocalization.[11]
Three remedies, in order of cost. Narrow or shift the detection band away from the overlap; this sacrifices some signal. Scan sequentially, one laser at a time, so the green dye is barely excited while the red channel records; this doubles the acquisition time. Unmix: record each dye’s contribution to each channel from single-labeled controls and solve the linear system pixel by pixel. Unmixing assumes the signal is a weighted sum of the pure spectra measured under the same settings.[11]
Spectral detectors and tunable lasers let you place both the excitation lines and the detection bands by wavelength, which makes the first two remedies far more flexible than fixed filter sets.[11]
Bands, lasers, three cells and the mixing matrix
Set the two detection bands and the two laser lines (the 405 or 448 nm diode, or any line from 485 to 790 nm from the white light laser, as on the Core’s Leica Stellaris), then compare simultaneous with sequential scanning. The cells show what each detector records and what linear unmixing recovers; the matrix is computed from the measured spectra.
Simultaneous: 3.0% of EGFP’s green-channel signal lands in the red channel, so the EGFP-only cell shows up faintly in the red channel and the merge gives it a faint false second color. In the double-labeled cell, 10% of the red channel is EGFP at 1.0× mCherry per EGFP. Unmixing removes it with almost no noise amplification (1.00×), because the two signatures are far apart (88°).
Spectra: FPbase[21], EGFP and mCherry, sampled every 5 nm from 380 to 750 nm, with the exact values at 405, 488, 561 and 640 nm; in between, a monotone cubic through the samples (no overshoot). Signal of a dye in a channel = brightness ε·Φ (EGFP 55,900 × 0.60, mCherry 72,000 × 0.22) × its relative absorption summed over the lasers on during that channel’s scan (both lasers delivering equal photon flux; at equal power the longer line delivers more photons, 15 % more at 561 than at 488 nm) × the share of its emission inside the band (trapezoid at 1 nm). EGFP’s measured emission spectrum ends at 700 nm, so the red band stops there, as in lesson 7. The measured excitation spectra end at 700 nm (EGFP) and 650 nm (mCherry): a laser beyond them reads “no data” and the figures count it as exciting nothing. mCherry’s measured emission starts at 550 nm, so its spill into a green band that ends there, or any spill below 0.01 %, reads “too little to measure”. The matrix is normalized to each dye’s own channel; the readouts are its off-diagonal terms, so brightness cancels in them. Noise after unmixing: length of the corresponding row of M−1, the factor by which equal, independent channel noise is amplified; the verdict also compares the photon noise of the unmixed mCherry in the double-labeled cell with the mCherry-only cell. Cells: equal numbers of EGFP molecules in the first and third cell, the chosen mCherry ratio in the second and third; 400 photons in the brightest pixel of the reference (EGFP, 488 nm, 500–550 nm), 3 photons of background, Poisson noise; unmixing applies M−1 pixel by pixel to the recorded counts. The two bands cannot overlap, as with a spectral detector or a dichroic splitter. Detector efficiency, laser power differences, reflected laser light, autofluorescence and maturation are not modeled.
The math, with its sources
About the model
Two fluorophores with measured spectra and their published brightness (ε·Φ), at the abundance ratio you set, with ideal detectors and the two lasers delivering equal numbers of photons per second (at equal power the 561 nm line would deliver 15 % more photons than the 488 nm line). Real crosstalk also depends on the relative brightness of the two labels: a weak red label next to a bright green one is the worst case.
At the microscope
Always image single-labeled controls at the final settings; they give the bleed-through coefficients and the unmixing references.
For colocalization measurements, use sequential scanning or unmixing, and report which.