Microscopy basics · Lesson 7 of 8 · about 5 minutes

When colors leak

Dye colors are broad. Part of a green dye’s light reaches the red channel and can look like a signal that isn’t there.

Trades: Speed · Clean image Widefield too Go deeper: the advanced module on spectra and bleed-through →
Green channel · 500–550 nm
Red channel · 600–650 nm
MergeSIMULATED · 22.5 × 14 µm
Green : red signal · brightest structures10 : 1
Lasers while the red channel records488 + 561 nm, both on
Bleed-through · of EGFP’s green signal3.0%
Ghost on the brightest microtubulesabout 30% of the nucleus
SIMULATED · A cell with green microtubules (EGFP) and a red nucleus (mCherry), both lasers on, green signal 10× the red (brightest structures). The microtubules also appear in the red channel, at about 30% of the nucleus’s brightness where they are brightest, around the centrosome: bleed-through.
1 · The idea

Emission curves have long tails

Dyes don’t give off a single color. Each emits a band of colors, with a long tail toward the red. When you image two dyes, part of the green dye’s light lands in the window meant for the red one. This is bleed-through, or crosstalk.

It matters most when you ask whether two things are in the same place (colocalization): green bleeding into the red channel makes everything green look a little red too.

Because the tails run toward longer wavelengths, green leaks into red far more than red leaks into green. Two fixes: choose dyes whose colors are far apart, and record the channels one after the other (sequential scanning), so the red channel is recorded while only the red dye’s laser is on.

COMPUTED · FPBASE SPECTRA green channel500–550 nmred channel600–650 nm 40045050055060065070075000.51wavelength (nm)relative emission (each dye’s own peak) EGFPmCherry mCherry in the greenband: below 0.01% ofits red-band light(too little to measure) × 10 · RED BAND EGFP × 10mCherry 3.0%of EGFP’s green-band light
COMPUTED · Measured emission spectra (FPbase), each scaled to its own peak, with an approximate screen colour for each wavelength under the axis (CIE 1931 → sRGB; pure spectral colours lie outside what screens can show). The bright sliver is EGFP light inside the red band; the inset magnifies it ten times. The measured spectra end at 700 nm (EGFP) and 750 nm (mCherry); mCherry still emits 17% of its peak at 700 nm and 5.4% at 750 nm, so its tail continues beyond the plot.
2 · Try it

Move the red window

A cell with EGFP on its microtubules and mCherry in its nucleus, with the two proteins’ measured spectra. Move the band the red channel collects, choose whether both lasers are on together or one after the other, and set how much stronger the green signal is than the red one (brightest structures).

Measured spectra (FPbase), each scaled to its own peak: emission filled, excitation dashed, laser lines in their own colour. Bright fill: EGFP light inside the red band.
What the two detectors record: the cell with EGFP on its microtubules and mCherry in its nucleus. Photon noise included; each channel shown on its own brightest real structure.
Every 50 nm red band at once: your setting on the curves of all settings. 03%6%9% 020%40%60% 570590610630650 red band starts at (nm) · each band 50 nm wide EGFP leak (left axis)mCherry light (right axis) most mCherry: 590–640 COMPUTED
Red band from600–650 nm
Presets
Lasers · while the red channel records
Green signal, × the red signal10×
Brightest structure in each channel, 600–650 nm band, both lasers on. Equal numbers of molecules give ≈ 3.8× (EGFP is brighter and better matched to its laser and band).
Presets
EGFP light in the red channel · of its green signal3.0%
mCherry light collected · share of its emission55%
False red signal where the microtubules are brightest · of the real red signal30% green signal 10× the red one: a faint false double label

Faint false signal: the microtubules appear in the red channel at 30% of the real red signal. Easy to miss by eye, but it biases any colocalization measurement. Moving further right cuts the leak but also mCherry light (now 55% of its emission).

SIMULATED / COMPUTED · Spectra: FPbase EGFP and mCherry (fpbase.org), sampled every 5 nm and interpolated with a monotone cubic, the same computation as module 07, so the two pages agree at every band. Bleed-through = EGFP light inside the red band ÷ EGFP light inside the green band (500–550 nm); mCherry’s share is of its emission measured to 750 nm. Equal detector efficiency at all wavelengths is assumed (real GaAsP and hybrid detectors are less sensitive in the red, which changes the measured value but not the logic). Signals use each protein’s brightness (extinction coefficient × quantum yield: EGFP 55,900 × 0.60, mCherry 72,000 × 0.22) and how strongly the lasers that are on excite it; both lasers deliver the same number of photons per second (at equal power the 561 nm beam carries 15% more photons). “One after the other”: the red channel is recorded with only the 561 nm laser, which excites EGFP at 0.43% of its 488 nm level; with both lasers on, the 488 nm line also excites mCherry, at 12% of its 561 nm level (included). The red band can be moved up to 650–700 nm because EGFP’s measured emission spectrum ends at 700 nm; redder bands would undercount its leak. Cell: the site’s simulated cell with EGFP on the microtubules and mCherry in the nucleus (confocal, 1 Airy unit, NA 1.4 oil; the nucleus imaged at 561/610 nm), 22.5 × 14 µm, 90 nm pixels; the red channel’s EGFP light reuses the 520 nm image (this changes sharpness only, not the leak). Brightness ratio = the brightest structures of each channel (99.5th percentiles: the centrosome region, where microtubules meet, in green; the nucleus in red) with both lasers on and the 600–650 nm band; 10× ≈ 2.6 EGFP molecules per mCherry. Single microtubule bundles away from the centrosome are at about 60% of that level, and so is their ghost. 400 photons in the brightest green structures at the reference setting, 2 photons background, photon noise; each channel is displayed on its own brightest real structure. Autofluorescence and detector noise are not modelled. On colocalization and bleed-through: Bolte & Cordelières 2006, J Microsc 224:213.

Remember

  1. 1Emission bands are broad, with long red tails: green leaks into red far more than red into green.
  2. 2Pick dyes whose colors are far apart, and detection bands that avoid the other dye’s light.
  3. 3Sequential scanning removes most bleed-through, at the cost of time.
Check yourself

A sample stained only with a green dye shows a faint copy of the green structures in the red channel. What is it?

Training

These pages are purely to improve your knowledge: practical training on the instrument is not only recommended but mandatory.