When colors leak
Dye colors are broad. Part of a green dye’s light reaches the red channel and can look like a signal that isn’t there.
Emission curves have long tails
Dyes don’t give off a single color. Each emits a band of colors, with a long tail toward the red. When you image two dyes, part of the green dye’s light lands in the window meant for the red one. This is bleed-through, or crosstalk.
It matters most when you ask whether two things are in the same place (colocalization): green bleeding into the red channel makes everything green look a little red too.
Because the tails run toward longer wavelengths, green leaks into red far more than red leaks into green. Two fixes: choose dyes whose colors are far apart, and record the channels one after the other (sequential scanning), so the red channel is recorded while only the red dye’s laser is on.
Move the red window
A cell with EGFP on its microtubules and mCherry in its nucleus, with the two proteins’ measured spectra. Move the band the red channel collects, choose whether both lasers are on together or one after the other, and set how much stronger the green signal is than the red one (brightest structures).
Faint false signal: the microtubules appear in the red channel at 30% of the real red signal. Easy to miss by eye, but it biases any colocalization measurement. Moving further right cuts the leak but also mCherry light (now 55% of its emission).
SIMULATED / COMPUTED · Spectra: FPbase EGFP and mCherry (fpbase.org), sampled every 5 nm and interpolated with a monotone cubic, the same computation as module 07, so the two pages agree at every band. Bleed-through = EGFP light inside the red band ÷ EGFP light inside the green band (500–550 nm); mCherry’s share is of its emission measured to 750 nm. Equal detector efficiency at all wavelengths is assumed (real GaAsP and hybrid detectors are less sensitive in the red, which changes the measured value but not the logic). Signals use each protein’s brightness (extinction coefficient × quantum yield: EGFP 55,900 × 0.60, mCherry 72,000 × 0.22) and how strongly the lasers that are on excite it; both lasers deliver the same number of photons per second (at equal power the 561 nm beam carries 15% more photons). “One after the other”: the red channel is recorded with only the 561 nm laser, which excites EGFP at 0.43% of its 488 nm level; with both lasers on, the 488 nm line also excites mCherry, at 12% of its 561 nm level (included). The red band can be moved up to 650–700 nm because EGFP’s measured emission spectrum ends at 700 nm; redder bands would undercount its leak. Cell: the site’s simulated cell with EGFP on the microtubules and mCherry in the nucleus (confocal, 1 Airy unit, NA 1.4 oil; the nucleus imaged at 561/610 nm), 22.5 × 14 µm, 90 nm pixels; the red channel’s EGFP light reuses the 520 nm image (this changes sharpness only, not the leak). Brightness ratio = the brightest structures of each channel (99.5th percentiles: the centrosome region, where microtubules meet, in green; the nucleus in red) with both lasers on and the 600–650 nm band; 10× ≈ 2.6 EGFP molecules per mCherry. Single microtubule bundles away from the centrosome are at about 60% of that level, and so is their ghost. 400 photons in the brightest green structures at the reference setting, 2 photons background, photon noise; each channel is displayed on its own brightest real structure. Autofluorescence and detector noise are not modelled. On colocalization and bleed-through: Bolte & Cordelières 2006, J Microsc 224:213.
Remember
- 1Emission bands are broad, with long red tails: green leaks into red far more than red into green.
- 2Pick dyes whose colors are far apart, and detection bands that avoid the other dye’s light.
- 3Sequential scanning removes most bleed-through, at the cost of time.