Slicing in 3D
Each confocal image is a thin slice. Step the focus and image again, and a stack of slices becomes a 3D picture.
A confocal image is a slice
Because the pinhole blocks light from above and below the focus (lesson 3), each confocal image shows only a thin layer of the sample: an optical section. With a 1.4 NA oil objective and the pinhole at 1 AU, it is about half a micrometer thick.
Move the focus up a little, take another image, and repeat. The result is a z-stack: a pile of slices that software can show in 3D, rotate, or project into a single picture.
How far apart should the slices be? Close enough that nothing falls between them: roughly half the section thickness, about 0.25 µm here. Larger steps are faster and gentler but can miss small things; much smaller steps add time and light without adding information.
Step through a stack
A simulated cell imaged as a confocal z-stack. Choose the step between slices, then move through the stack. The side view shows what your slices capture, and what falls between them.
About right: two slices per section, so nothing falls between them. The side view rebuilt from your slices holds everything the optics deliver, in blocks half a section tall.
SIMULATED · The same simulated cell as in lessons 3 and 5, computed as 37 confocal slices from 0 to 4.5 µm above the coverslip, 0.125 µm apart (1 Airy unit, 1.4 NA oil objective, 488 nm laser, 520 nm light, vectorial point spread function, sample index-matched to the oil). “The optics deliver” blends those fine slices; “your stack” shows each of your slices over its own step, as many viewers do, with photon noise (about 100 photons per pixel in the brightest microtubules and DNA). The bells on the right are each slice’s response to a thin glowing layer: 501 nm thick at half height (a single point gives 391 nm). “About right” follows the practical rule: a step of half the section. The strict rule used for deconvolution is about 0.13 µm (module 06). Light: every slice lights the whole depth of the cell, not only the slice being recorded.
Remember
- 1Each confocal image is an optical section, about 0.5 µm thick with a 1.4 NA oil objective at 1 AU.
- 2A z-stack is many sections, about half a section thickness apart.
- 3Every slice costs time and light, and the laser fades the whole depth each time, not just the slice.