Microscopy basics · Lesson 6 of 8 · about 5 minutes

Slicing in 3D

Each confocal image is a thin slice. Step the focus and image again, and a stack of slices becomes a 3D picture.

Trades: Detail · Speed · Sample health Widefield too Go deeper: the advanced module on z-stacks →
Simulated z-stack: five confocal sections of one cell stacked in perspective, from the microtubules near the coverslip to the top of the nucleus SIMULATED
Five confocal sections through one simulated cell, from the microtubules near the coverslip to the top of the nucleus.
1 · The idea

A confocal image is a slice

Because the pinhole blocks light from above and below the focus (lesson 3), each confocal image shows only a thin layer of the sample: an optical section. With a 1.4 NA oil objective and the pinhole at 1 AU, it is about half a micrometer thick.

Move the focus up a little, take another image, and repeat. The result is a z-stack: a pile of slices that software can show in 3D, rotate, or project into a single picture.

How far apart should the slices be? Close enough that nothing falls between them: roughly half the section thickness, about 0.25 µm here. Larger steps are faster and gentler but can miss small things; much smaller steps add time and light without adding information.

z-stepcoverslipeach band: one optical section
Schematic side view, not to scale. At the recommended step the sections overlap: see it to scale below.
2 · Try it

Step through a stack

A simulated cell imaged as a confocal z-stack. Choose the step between slices, then move through the stack. The side view shows what your slices capture, and what falls between them.

Top view: the slice you are on. Drag up or down to move the orange line.
Side view along the orange line, to scale. Drag the divider to compare.
Simulated side view of the cell along the orange line, 13.0 µm wide and 4.75 µm tall, to scale. Left of the divider, what the optics deliver; right of it, the stack rebuilt from 19 slices 0.25 µm apart, each shown over its own step. Sections are 0.5 µm thick. Use the left and right arrow keys to move the divider.
your sliceseach slice’s optical section, 0.5 µm thickmicrotubulesDNA
Step between slices0.25 µm
Slice · height above the coverslip3 of 19 · 0.5 µm
Step, for a 1.4 NA oil objective · sections 0.5 µm thick oversampled about right undersampled 0.10.250.512 µm 0.25 µm
Microtubules on the coverslip · brightness at each height 01234 height above the coverslip (µm) best slice: 100%
Slices for this cell · 0 to 4.5 µm19
Slices per section2
Time · 1 s per slice19 s
Light on the cell · vs one slice19×

About right: two slices per section, so nothing falls between them. The side view rebuilt from your slices holds everything the optics deliver, in blocks half a section tall.

SIMULATED · The same simulated cell as in lessons 3 and 5, computed as 37 confocal slices from 0 to 4.5 µm above the coverslip, 0.125 µm apart (1 Airy unit, 1.4 NA oil objective, 488 nm laser, 520 nm light, vectorial point spread function, sample index-matched to the oil). “The optics deliver” blends those fine slices; “your stack” shows each of your slices over its own step, as many viewers do, with photon noise (about 100 photons per pixel in the brightest microtubules and DNA). The bells on the right are each slice’s response to a thin glowing layer: 501 nm thick at half height (a single point gives 391 nm). “About right” follows the practical rule: a step of half the section. The strict rule used for deconvolution is about 0.13 µm (module 06). Light: every slice lights the whole depth of the cell, not only the slice being recorded.

Remember

  1. 1Each confocal image is an optical section, about 0.5 µm thick with a 1.4 NA oil objective at 1 AU.
  2. 2A z-stack is many sections, about half a section thickness apart.
  3. 3Every slice costs time and light, and the laser fades the whole depth each time, not just the slice.
Check yourself

Your optical sections are about 0.5 µm thick. Which z-step avoids gaps without wasting light?